24 | Chapter 2: Installation
Without Buffer Switching (1 Buffer)
1. Place the supply bottle(s) on the same
horizontal plane as the instrument.
2. Connect one the tubes to the
port.
Fluid In

Final Check

Verify that the tubing was not crimped during installation.
l
Ensure that there are no loose fittings or cable connections.
l

Attach the Mist Shield

1. Position the mist shield so the
gaps align with the
thumbscrews. The top rests on
the two rubber pads above the
manifold.
Always lift the mist shield straight
up, not towards you, when
removing it.
405™ Microplate Washer LS
Dispense