Remove/Change Wash Manifold; Replace The Vacuum Pump Fuse - Biotek 405 LS Operator's Manual

Microplate washer
Table of Contents

Advertisement

Remove/Change Wash Manifold

To clean or replace the wash manifold:
1. Run the system "dry" to remove any fluid: connect an empty supply bottle and
run a prime protocol until the tubing is empty.
2. Turn off the instrument, disconnect the power cable, and remove the mist
shield.
3. Release the thumbscrews that secure the manifold(s) in place.
Dual manifolds: Hold the two manifolds together as a single unit when
n
removing and reinstalling.
4. Carefully slide the manifold(s) and end plates off their rails.
DO NOT AUTOCLAVE
n
Reverse these steps to reinstall the manifold. Be sure to correctly position the
manifold(s) on the rails; they only fit one way: the side with the recessed channel
openings faces the washer. Make sure there are no gaps between the manifold and
the washer.
Change the Instrument Settings
When you change the type of manifold, e.g. from a 96-tube to a 192-tube manifold,
you must tell the instrument which manifold is installed:
LHC
1. Select
Tools>Instrument
Utilities>Washer
2. Update the
Manifold Selection
3. Click
.
Send

Replace the Vacuum Pump Fuse

Spare fuses (PN 46055) are shipped with the instrument in case the pump blows a
fuse.
Tools: Screwdriver
To change the fuse:
the manifold!
Keypad
1. Press
Setup Menu
g
2. Select
and then
3. Select the currently installed wash manifold by
pressing its Soft-key: Single or Dual 96-tubeor 192-tube
.
manifold.
Replace Components | 145
(button in center of keypad).
.
MANIF
BioTek Instruments, Inc.

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 405 LS and is the answer not in the manual?

Table of Contents

Save PDF