Liquid Testing the 405™ Microplate Washer LS

Which Tests to Perform?

We recommend that you perform these routine tests before first use (after the
IQ) and then monthly:
Washer
Dispense Precision Test.
l
from tube to tube across the manifold.
Evacuation Efficiency Test.
l
aspiration aspect of plate washing. The lower the residuals per well, the better the
evacuation efficiency of the washer.
Specially equipped washers: with Buffer Switching and/or Verify™ Technology,
perform test annually (starting one year after installation):
Annual Buffer Switching Test.
l
tube to tube across the manifold, for each valve, when the washer is equipped with
buffer switching.
Verify Sensor Test.
l
against a microplate absorbance reader's determination to ensure accurate, repeatable
volume-level detection.
Perform the tests applicable to the installed manifold type and feature set:
If you have:
96-Tube Manifold
192-Tube Manifold
Buffer Switching
Vacuum Filtration Module
Verify™ Technology
405™ Microplate Washer LS
Precision tests measure the variability of volumes dispensed
Measures the residual volume per well after the
Measures the variability of volumes dispensed from
Compares the ultrasonic level sensor's measurement precision
Run Liquid Test(s):
Dispense Precision Test (96)
Evacuation Efficiency Test (96)
Dispense Precision Test (192)
Evacuation Efficiency Test (192)
Buffer Switching Dispense Test
Vacuum Filtration Evacuation Efficiency (for either
96- or 384-well plates)
Verify Sensor Test