The Labnet 311DS Shaking Incubator comes with one removable full shelf and four large, rubber padded adjustable feet for leveling the unit. Additional half and full shelves may be purchased. The 311DS shaking mechanism supports a wide range of optional rack and flask platforms which are user changeable.
4.0 Unpacking Upon receipt of your Labnet 311DS Shaking Incubator, examine the carton and unit for damages. If shipping damage has occurred, a claim must be filed with the carrier. The carrier is responsible for correcting shipping damages. Save all packaging until the unit has been shown to operate properly to your satisfaction.
6.0 Controls The controls for the Labnet 311DS Shaking Incubator include power to the oven, temperature setting, safety temperature overshoot setting, shaker speed, and shaking time. In addition there is a door-open shaker shut-off switch and a chamber light switch.
7.0 Calibration The Labnet 311DS Shaking Incubator is calibrated at 37°C at the factory. The unit can be recalibrated after the chamber temperature has stabilized at the setpoint for several hours. Suspend a certified reference thermometer in the chamber. Compare the units display to the reference thermometer.
. Should you have a question about the Labnet 311DS Shaking Incubator or require service for the unit, contact Corning Customer service at 800.492.1110 or 978.442.2200. Do not send a unit for service without first calling to obtain a repair authorization (RA) number. The unit should be properly packed to avoid damage.
12.0 Limited Warranty Corning Incorporated (Corning) warrants that this product will be free from defects in material and workmanship for a period of one (1) year from date of purchase. CORNING DISCLAIMS ALL OTHER WARRANTIES WHETHER EXPRESSED OR IMPLIED, INCLUDING ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR OF FITNESS FOR A PARTICULAR PURPOSE.