Labnet I5311-DS Instruction Manual

Incubator shaker

Advertisement

Quick Links

Labnet Incubator Shaker
I5311-DS and I-5311-DS-230V
Instruction Manual
PO Box 841 Woodbridge, NJ 07095 USA
Phone: 732-417-0700
Fax: 732-417-1750
www.labnetlink.com
Copyright July 2004

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the I5311-DS and is the answer not in the manual?

Questions and answers

Summary of Contents for Labnet I5311-DS

  • Page 1 Labnet Incubator Shaker I5311-DS and I-5311-DS-230V Instruction Manual PO Box 841 Woodbridge, NJ 07095 USA Phone: 732-417-0700 Fax: 732-417-1750 www.labnetlink.com Copyright July 2004...
  • Page 2: Table Of Contents

    Accessories page 7 GRAPHIC SYMBOLS Your Labnet I-5311-DS incubator shaker uses internationally accepted graphic symbols to help convey information to the user and to call the users attention to important safety precau- tions and guides for using this equipment. indicates that user should consult manual further description or discussion.
  • Page 3: Introduction

    INTRODUCTION This manual covers the specifications, operation and use of the Labnet I-5311-DS shaking incuba- tor and its accessories. Please pay special attention to the Safety Precautions section in this manual. The I-5311-DS shaking incubator from Labnet International provides an extremely stable tempera- ture environment, and an aggressive shaking motion for the mixing of materials for cell culture growth at ambient and above ambient temperatures.
  • Page 4: Installation

    CONTROLS & CALIBRATION The controls for the I5311-DS Shaker Incubator include power to the oven, temperature setting, safety temperature overshoot setting and shaker speed and shaking time. In addition there is a door-open shaker shut-off switch and a chamber light switch.
  • Page 5 -Heating Indicator-This indicator will illuminate GREEN when the controller is calling for heat from the heater. This indicator will be on continuously while the oven heats up to the set tem- perature and will then cycle on and off at the set temperature. -Safety Thermostat-The safety thermostat located on the back of the unit is manually set and completely independant of the Main Termperature Controller.
  • Page 6: Operating The I-5311-Ds

    OPERATING THE LABNET I-5311-DS INCUBATOR SHAKER As advised in the installation section, it is very important that the unit be level and placed on a very solid surface. When loading the platforms, it is also important to load these in a balanced manner or the unit can vibrate and possibly walk.
  • Page 7: Troubleshooting Guide

    Department at 732-417-0700 between the hours of 9:00am and 5:00pm EST. If service necessitates the return of the unit to Labnet’s Service Department, a return authorization (RA) number must first be obtained and a decontamination certificate completed. The unit must be thoroughly decontaminated before shipping.
  • Page 8: Accessories

    Flask clamps are available in 50ml, 125ml, 150ml, 500ml, 1000ml, and 2000ml. Test tube rack clamps are also available. Labnet is continuously adding accessories. Check with your representative or call Labnet directly at 732-417-0700 or check our website: www.labnetlink.com for a complete listing of accessories...