Section 5 Troubleshooting Guide
Many operating problems may be solved by carefully reading and following the
instructions in this manual accordingly. Some suggestions for troubleshooting
are given below. Should these suggestions not resolve the problem, please
contact our SERVICE DEPARTMENT or a distributor in your region for
assistance. If troubleshooting service is required, please include a full
description of the problem.
Problem
Screen does
No Signal from the CMOS
camera
Light lamp does
UV light doesn t light up
Cannot print the image file
on 64 bit system (e.g. Select
the printer type as Adobe
PDF).
Cannot operate the lights, or
capture the images.
Suggestion
Check the main power switch is on
Please check if the camera is on.
Check power input cable between CMOS camera
and chamber is connected well.
Check video output cable between CMOS camera
and chamber is connected well.
up
Check white light lamp is switched on
Check UV transilluminator is switched on.
Since 312nm wavelength UV is invisible light, you
can put on a sheet of white paper onto the
transilluminator to help distinguish if the light is
on.
A4
Paper
Light OFF
Please select the PDF Creator type to print the
image.
PDF Creator download web site:
http://www.pdfforge.org/pdfcreator
Check if the device is connected with your
computer.
If the device is connected with the computer, then
check if Bluetooth function in the computer is off.
(*Note1)
88
A4
Paper
Light ON