Advertisement

Quick Links

Coronis Fusion 4MP / 6MP
User Guide
MDCC-4230, MDCC-6330
K5902056/00
31/10/2014

Advertisement

Table of Contents
loading

Summary of Contents for Barco MDCC-4230

  • Page 1 Coronis Fusion 4MP / 6MP User Guide MDCC-4230, MDCC-6330 K5902056/00 31/10/2014...
  • Page 2 Barco nv President Kennedypark 35, 8500 Kortrijk, Belgium Phone: +32 56.23.32.11 Fax: +32 56.26.22.62 Support: www.barco.com/esupport Visit us at the web: www.barco.com Printed in Belgium...
  • Page 3: Table Of Contents

    Table of contents TABLE OF CONTENTS 1. Welcome! ..................3 About the product .................... . 3 What’s in the box.
  • Page 4 Table of contents 6. Cleaning your display ................35 Cleaning instructions ..................35 7.
  • Page 5: Welcome

    • an external power supply If you ordered a Barco display controller, it’s also in the box together with its accessories. A dedicated user guide is available on the documentation CD. Keep your original packaging. It is designed for this display and is the ideal protection during transport and storage.
  • Page 6 1. Welcome! K5902056 CORONIS FUSION 4MP / 6MP 31/10/2014...
  • Page 7: Parts, Controls And Connectors

    2. Parts, controls and connectors 2. PARTS, CONTROLS AND CONNECTORS Display front view Overview Image 2-1 Front view Left key USB downstream connector Right key Menu key Standby key Power LEDs Ambient light sensor The key icons are displayed above the keys, adapted to the function that it is used for (menu dependent).
  • Page 8: Display Rear View

    2. Parts, controls and connectors Display rear view Overview Image 2-2 Rear view Connector compartment cover +24 VDC power input Neck cover Display stand cover DisplayPort 2 video input DVI 2 video input DisplayPort 1 video input DVI 1 video input USB upstream connector USB downstream connectors K5902056 CORONIS FUSION 4MP / 6MP 31/10/2014...
  • Page 9: Display Installation

    Prior to installing your Coronis Fusion 4MP / 6MP and connecting all necessary cables, make sure to have a suitable display controller physically installed in your computer. If you are using a Barco display controller, please consult the user guide delivered with it to do this.
  • Page 10: Unlocking The Height Mechanism

    3. Display installation 2. Slide the cover upwards while holding the clips pressed. Unlocking the height mechanism In the factory, the height-positioning system in the display stand is locked with a red hook to prevent damage during transportation. You’ll have to remove this hook before adjusting your display height position.
  • Page 11: Adjusting The Display Position

    3. Display installation 3. Keep the clip in the dedicated hole in case the display needs to be shipped later. To retain the hook for possible future transportation, insert the short, red end of the hook back into the stand of your display. Adjusting the display position Now that the height-positioning system of your display is unlocked, you can adjust the display position.
  • Page 12: Connecting The Signal Cables

    3. Display installation Do not try to pivot your display when attached to the stand. Trying to do so AUTION could cause serious damage to your display and its stand. Connecting the signal cables To get access to the connectors, remove the connector compartment cover. See "Re- moving the covers", page 7 .
  • Page 13: Connecting The Power Cable

    3. Display installation Connecting the power cable To connect the power cable to the display: 1. Connect the supplied external DC power supply to the +24 VDC power input of your Coronis Fusion 4MP / 6MP display. 2. Plug the other end of the external DC power supply into a grounded power outlet by means of the proper power cord delivered in the packaging.
  • Page 14: Re-Attaching The Covers

    3. Display installation Tip: The cable straps at the inside of the connector compartment allow you to fix the cables for better shielding of the cables. Re-attaching the covers To re-attach the display stand cover 1. Slide the display stand cover downward. You’ll hear a “click” sound of the cover’s clips when the display stand cover is in position.
  • Page 15: Vesa-Mount Installation

    3. Display installation VESA-mount installation Use an arm that is approved by VESA (according to the VESA 100 mm stan- ARNING dard). Use an arm that can support the weight of the display. Refer to the technical specifica- tions of this display for the applicable weight. AUTION You should mount the panel in landscape position.
  • Page 16: First Time Starting Up

    3. Display installation 2. Slide the cover upwards while holding the lower sides pulled open. 3. Lift the plastic frame that covers the fixation of the panel to the stand. Turn it for 45 degrees to uncover the fixation screws. 4.
  • Page 17 Your Coronis Fusion 4MP / 6MP display will be running in a basic video mode at a de- fault refresh rate when first time starting up. If you are using a Barco display controller, please consult the dedicated user guide available on the system CD to install the drivers, software and documentation.
  • Page 18 3. Display installation K5902056 CORONIS FUSION 4MP / 6MP 31/10/2014...
  • Page 19: Daily Operation

    LCD displays use technology based on pixels. As a normal tolerance in the manufacturing of the LCD, a limited number of these pixels may remain either dark or permanently lit, without affecting the diagnostic performance of the product. To ensure optimal product quality, Barco applies strict selection criteria for its LCD panels.
  • Page 20: Key Indicator Lights

    The ’MediCal QAWeb’ system offers online service for high-grade Quality Assurance, providing maximum diagnostic confidence and uptime. Barco recommends to install MediCal QAWeb Agent and apply the default QAWeb policy at least. This policy includes calibration on regular intervals. Connecting to MediCal QAWeb Server offers even more possibilities.
  • Page 21: Bringing Up The Osd Menus

    4. Daily operation Bringing up the OSD menus How to bring up the OSD menus The OSD menu allows you to configure different settings to make your Coronis Fusion 4MP / 6MP fit your needs within your working environment. Also, you can retrieve general information about your display and its current configuration settings through the OSD menu.
  • Page 22 4. Daily operation Cancel Standby (IEC 60417–5009) K5902056 CORONIS FUSION 4MP / 6MP 31/10/2014...
  • Page 23: Advanced Operation

    5. Advanced operation 5. ADVANCED OPERATION OSD menu language About the OSD menu language By default, the OSD menu comes up in English. However, there’s a wide range of other languages avail- able for the OSD menu of your Coronis Fusion 4MP / 6MP. To change the language of the OSD menu: 1.
  • Page 24: Key Indicator Lights

    By default, DPMS mode is enabled on your display, but it also needs to be activated on your workstation. To do this, go to the “Power options properties” window of your workstation. Barco recommends setting DPMS activation after 20 minutes of non-usage. When DPMS mode is enabled on your display, an additional OSD power saving function becomes available: hibernate.
  • Page 25: Hibernate

    5. Advanced operation 4. Select Enabled/Disabled as desired and confirm. Hibernate About hibernate Enabling hibernation will not only switch off the backlight but will also force the display to disable other functionalities so that power consumption is further reduced to a minimum. This happens after a specific period of time which can be manually adjusted.
  • Page 26: Color Presets

    5. Advanced operation Color presets About color presets The available color preset settings for your display are: Clearbase Simulation of the clearbase film color temperature. Bluebase Simulation of the bluebase film color temperature. User When selecting the User color temperature setting, you will be able to manually define the X and Y coordinates or the display color temperature in separate submenus.
  • Page 27: Viewing Modes

    5. Advanced operation To change the color coordinates: 1. Bring up the OSD main menu. 2. Navigate to the Configuration > Calibration > Color Settings menu. 3. Enter the Color Definition submenu. 4. Select Color Coordinates and confirm. 5. Enter the x and/or y submenu. 6.
  • Page 28: Ambient Light Compensation (Alc)

    In short, the DICOM display function results in more visible grayscales in the images. Barco recommends selecting the DICOM display function for most medical viewing applications. User This display function will be automatically selected when display functions are defined by MediCal...
  • Page 29: Dicom Status Report

    5. Advanced operation 5.15 DICOM status report About DICOM status report Following information is available: DICOM Compliance Status (status Compliance status Shows if the current DICOM curve since last compliance check) is compliant or not. Maximum error Shows the maximum error of the current DICOM curve.
  • Page 30: Dicom Calibration

    5. Advanced operation 3. Select DICOM compliance check to start the compliance check. Warning:Pressing a key during the compliance check will abort the check. 5.17 DICOM calibration About DICOM calibration The DICOM calibration will add a correction to the current DICOM curve to approach the perfect DICOM curve as good as possible.
  • Page 31: Reading Rooms

    5. Advanced operation 5.20 Reading rooms About reading rooms Reading rooms can only be selected when the DICOM display function is selected. Therefore, please refer to "Display functions", page 25 to correctly set the display function. The American Association of Physicists in Medicine (AAPM) composed a list of pre-defined reading rooms. Each of these reading rooms are defined by following parameters: •...
  • Page 32: Image Scaling

    5. Advanced operation 2. Navigate to the Configuration > Calibration > Ambient Light menu. 3. Enter the Continuous ALC submenu. 4. Select Enabled/Disabled as desired and confirm. 5.22 Image scaling About image scaling Enabling image scaling will multiply each individual pixel to one or more adjacent pixels so that the size of the displayed image will be a multiple of the original image source video input signal.
  • Page 33: Video Input Signals

    5. Advanced operation Two Image Sources This mode is intended for displaying and manually configuring two connected video input signals (one on each display side). When selecting this mode, the video settings become available for the selected video input signal on each side of the display. Expert mode This mode is intended for displaying and manually configuring one or two connected video input...
  • Page 34: Grayscale Conversion Modes

    5. Advanced operation 2. Navigate to the Configuration > Image Sources menu. 3. Enter one of the available submenus as desired. 4. Select one of the available connectors/inputs and confirm. 5.25 Grayscale conversion modes Your Coronis Fusion 4MP / 6MP display automatically detects the connected video in- put signals and applies the correct grayscale conversion settings.
  • Page 35: Edid Timings

    5. Advanced operation 5.27 EDID timings About EDID timings Following EDID timings are available for your Coronis Fusion 4MP / 6MP: Resolution Allows to manually modify the resolution of the image source video input signal. Refresh Rate Allows to manually select the refresh rate of the image source video input signal depending on the maximum refresh rate of the display controller connected to your display.
  • Page 36 5. Advanced operation K5902056 CORONIS FUSION 4MP / 6MP 31/10/2014...
  • Page 37: Cleaning Your Display

    6. Cleaning your display 6. CLEANING YOUR DISPLAY Cleaning instructions To clean the display Clean the display using a sponge, cleaning cloth or soft tissue, lightly moistened with a recognized clean- ing product for medical equipment. Read and follow all label instructions on the cleaning product. In case of doubt about a certain cleaning product, use plain water.
  • Page 38 6. Cleaning your display K5902056 CORONIS FUSION 4MP / 6MP 31/10/2014...
  • Page 39: Repackaging Instructions

    7. Repackaging instructions 7. REPACKAGING INSTRUCTIONS Overview of the packaging Image 7-1 Repack instructions Coronis Fusion 4MP / 6MP display System CD, Documentation CD Top buffer and User guide Display controller box Bottom buffer Power supply Cables (video, USB, power) K5902056 CORONIS FUSION 4MP / 6MP 31/10/2014...
  • Page 40: Repacking Your Display

    7. Repackaging instructions Repacking your display How to repack your display 1. Place the empty box on a stable surface. 2. Slide the system CD, documentation CD and user guide in the dedicated bag attached at the inside of the box. 3.
  • Page 41: Important Information

    8. Important information 8. IMPORTANT INFORMATION Safety information General recommendations Read the safety and operating instructions before operating the device. Retain safety and operating instructions for future reference. Adhere to all warnings on the device and in the operating instructions manual. Follow all instructions for operation and use.
  • Page 42 A11:2011, EN 60601-1-2:2007, CCC - GB9254-2008 + GB4943.1-2011 + GB17625.1-2012, KCC, VCCI, FCC class B, ICES-001 Level B, FDA 510(k), RoHS. MDCC-4230: CE0120 (MDD 93/42/EEC; A1:2007/47/EC class IIb product), CE - 2004/108/EC, IEC 60950-1:2005 + A1:2009, EN 60950-1:2006 + A1:2010 + A11:2009 + A12:2011 + A2:2013, IEC 60601-1:2005, ANSI/AAMI ES 60601-1:2005 + C1:2009 + A2:2010, CAN/CSA C22.2 No.
  • Page 43: Environmental Information

    (Also called RoHS of Chinese Mainland), the table below lists the names and contents of toxic and/or hazardous substances that Barco’s product may contain. The RoHS of Chinese Mainland is included in the MCV standard of the Ministry of Information Industry of China, in the section “Limit Requirements of toxic substances in Electronic Information Products”.
  • Page 44 “Electronic Information Products Pollution Control Labeling Standard” of Chinese Mainland, marked with the Environmental Friendly Use Period (EFUP) logo. The number inside the EFUP logo that Barco uses (please refer to the photo) is based on the “Standard of Electronic Information Products Environmental Friendly Use Period”...
  • Page 45: Regulatory Compliance Information

    Regulatory compliance information Indications for use The Coronis Fusion 4MP / 6MP (MDCC-4230, MDCC-6330) Medical Flat Panel Display System is in- tended to be used as a tool in displaying and viewing digital images (excluding digital mammography) for review and analysis by trained medical practitioners.
  • Page 46: Emc Notice

    8. Important information EMC notice General information No specific requirement on the use of external cables or other accessories except power supply. With the installation of the device, use only the delivered power supply or a spare part provided by the legal manufacturer.
  • Page 47 8. Important information Immunity test Electromagnetic IEC 60601 Compliance level environment – Test levels guidance Electrostatic discharge ± 6kV contact ± 6kV contact Floors should be wood, (ESD) concrete or ceramic tile. ± 8kV air ± 8kV air If floors are covered with IEC 61000-4-2 synthetic material, the relative humidity should...
  • Page 48 8. Important information Immunity test Electromagnetic IEC 60601 Compliance level environment – Test levels guidance d = 2.3√P 800 MHz to 2.5 Where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in...
  • Page 49: Explanation Of Symbols

    8. Important information Rated maximum output Separation distance according to frequency of transmitter power of transmitter 150kHz to 80MHz 80MHz to 800MHz 800MHz to 2.5GHz d=1.2√P d=1.2√P d=2.3√P 0.01 0.12 0.12 0.23 0.38 0.38 0.73 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. These guidelines may not apply in all situations.
  • Page 50 8. Important information Indicates the device is approved according to the CCC regulations Indicates the device is approved according to the VCCI regulations Indicates the device is approved according to the KC regulations Indicates the device is approved according to the BSMI regulations Indicates the device is approved according to the PSE regulations Indicates the USB connectors on the device Indicates the DisplayPort connectors on the device...
  • Page 51: Legal Disclaimer

    Barco software products are the property of Barco. They are distributed under copyright by Barco N.V. or Barco, Inc., for use only under the specific terms of a software license agreement between Barco N.V. or Barco Inc. and the licensee. No other use, duplication, or disclosure of a Barco software product, in any form, is authorized.
  • Page 52 8. Important information Active screen size (H x V) 654.360 x 408.576 mm (25.76 x 16.09”) Aspect ratio (H:V) 16:10 Resolution Native 6MP (3280 x 2048) Configurable to 2 x 3MP+ (1640 x 2048) Configurable to 2 x 3MP (1536 x 2048) Pixel pitch 0.1995 mm Color imaging...
  • Page 53: Technical Specifications

    Operating humidity 8% - 80% (non-condensing) Storage humidity 5% - 95% (non-condensing) Operation altitude 3000 m Technical specifications MDCC-4230 Product acronym MDCC-4230 Screen technology TFT AM Color LCD Dual Domain IPS-Pro Active screen size (diagonal) 756 mm (29.8") Active screen size (H x V) 641.28 x 400.8 mm (25.2 x 15.8")
  • Page 54 8. Important information Viewing angle (H, V) 178° Uniform Luminance Technology (ULT) Per Pixel Uniformity (PPU) Ambient Light Compensation (ALC) Backlight Output Stabilization Yes (2x) (BLOS) I-Guard Maximum luminance 750 cd/m² DICOM calibrated luminance 500 cd/m² (ULT off) Contrast ratio (typical) 1200:1 (panel typical) Response time (Tr + Tf) 20 ms...
  • Page 55: Open Source License Information

    Open source license usage This product contains software components released under an Open Source license. A copy of the source code is available on request by contacting your Barco customer support representative. EACH SEPARATE OPEN SOURCE SOFTWARE COMPONENT AND ANY RELATED DOCUMENTA- TION ARE PROVIDED "AS IS"...

This manual is also suitable for:

Coronis fusion 6mpCoronis fusion 4mpMdcc-6330