2.1.1 LOCATION AND LAYOUT

Mount the analyzer in a clean, vibration-free area avoiding direct radiant heat, sunlight
and drafts. Avoid areas containing chlorinating equipment. The 9033X Sodium analyzer
default design has the two main sections (transmitter and wet section) combined as a
single unit. If necessary the sensor unit can be mounted separately as long as it is no
more than 330 feet (100meters) from the transmitter unit.

2.1.2 TRANSMITTER UNIT

The transmitter unit controls the operations of the analyzer. Power supply, CPU card,
Analog card and input power terminal junction are housed in the transmitter enclosure.
Transmitter unit is a NEMA 4X rated enclosure with cable glands for wiring. Size and
layout of transmitter is shown below.
Figure 4. Dimensions of the transmitter unit. (All dimensions in inches)

2.2 SAMPLE REQUIREMENTS

The maximum sample pressures and temperatures specified in the SPECIFICATION
section should not be exceeded. The sample should be introduced to the system at a
temperature and pressure suitable for measurement. If necessary, customer may
choose to use sample cooling and pressure reducing equipment. It is highly
recommended that a flow meter is installed to the sample inlet stream to ensure that
the sample flow rate is within the specified range. When pressure reducing equipment
is being used, a pressure relief valve should be installed between the sample point and
sample inlet to ensure maximum safety.