8 SPECIFICATIONS

Range:
Accuracy:
Reproducibility:
Response Time:
Current Outputs:
Alarms:
Power:
Sample:
Ambient Temp:
Pressure:
Composition:
Power
Consumption:
Sample Inlet
Fitting:
Sample Outlet
Fitting:
Waltron User Manual 101-044-E.1
9033X Sodium Analyzer
0.1ppb – 10ppm
+/-5% of reading or +/- 0.1ppb (whichever is greater) within +/-
5C of calibration temperature
+/-5% of reading or +/- 0.1ppb (whichever is greater) at
constant temperature
90% of 1-10ppb step: less than 4 minutes; 90% of 100-1ppm
step: less than 6 minutes
Two isolated 4-20mA current (analog) outputs
Three voltage-free contacts, alarm points set from transmitter
rated at 2A, 250VAC (non-inductive)
Wide range of input power supply 90VAC – 250VAC
Temperature: 41-131F (5-55C); Flow 150-400ml/min
32-131F (0-55C)
5-30psig
Sample should be filtered to 60 microns, free of film forming
compounds
Less than 20VA
¼" Swagelok
Barbed fitting for 3/8" ID hose connection
53
9033X
53