Installation; Receiving; Analyzer Handling; Location And Mounting - Waltron 9096 User Manual

Degassed cation conductivity analyzer
Table of Contents

Advertisement

Installation

3 INSTALLATION

RECEIVING

3.1
The Waltron 9096 Desgassed Cation Conductivity Analyzer is assembled and fully tested for
proper performance. Before proceeding with analyzer installation, it is recommended to:
Check that the box and analyzer have not been damaged during transportation.
Take extreme care during analyzer unpacking and moving.
Be careful not to misplace accessories during unpacking. Refer to the included packing
list.

ANALYZER HANDLING

3.2
Take extreme care when lifting or moving the analyzer. Before moving the analyzer, it is
recommended to manually empty all of the liquid containing components.

LOCATION AND MOUNTING

3.3
It is recommended to install the analyzer in a suitable position:
The location is to be clean, covered and properly enclosed to provide the analyzer with
good ventilation and low dust concentration.
Operating environmental conditions are: temperature between 0 and 60°C (32-140°F) at
max 80% relative humidity. If the temperature is below 0°C, the analyzer should be
installed in a heated cabinet.
Place the analyzer close to the sample point to achieve the minimum response time; the
sample should be homogenous and representative.
The drain line should be properly dimensioned and positioned at a downward slope to
allow the drain of analyzed sample (by gravity).
WARNING!
15
The sample drain of the analyzer must be at ambient
pressure with no restriction or counter pressure.
Waltron User Manual 101-046-B.3
9096 Degassed Cation Conductivity Analyzer