3 INSTALLATION

RECEIVING

3.1
The Waltron 3049 Ethylene Glycol Analyzer is assembled and fully tested for proper
performance. Before proceeding with analyzer installation, it is recommended to:
Check that the box and analyzer have not been damaged during transportation.
Take extreme care during analyzer unpacking and moving.
Be careful not to misplace accessories during unpacking. Refer to the included packing
list.

ANALYZER HANDLING

3.2
Take care when lifting or moving the analyzer. Before moving the analyzer, it is recommended
to manually empty all of the hydraulic parts of any liquids.

LOCATION AND MOUNTING

3.3
It is recommended to install the analyzer in a suitable location::
The location is to be clean, covered and properly enclosed to provide the analyzer with
good ventilation and low dust concentration.
Required environmental conditions:
o
Because of chemicals and waste gases it is necessary to have a well ventilated location
for the analyzer.
The analyzer is supplied with four mounting brackets for wall or stainless steel support
rack installation. Use 4 screws M6 or M8 (1/4") to mount the analyzer. A panel mounting
kit is available from Waltron if desired.
Reagent bottles are supplied with the analyzer.
o A reagent shelf is provided with the instrument
o Allow a minimum of 14 inches (40 cm), but no more than 30 inches (75 cm),
The fast-loop reservoir should be mounted within reach of the sample tubing
(approximately 24" (60 cm) in length).

PRE-INSTALLATION

3.4
Before installing the analyzer, take the following precautions:
Place the analyzer close to the sample point to achieve the minimum response time; the
sample should be homogenous and representative.
The drain line should be properly dimensioned and positioned at a downward slope to
allow the drain of analyzed sample (by gravity) and the overflow coming from external
fast-loop reservoir (if used).
Waltron User Manual 101-042-B.1
3049 Ethylene Glycol Analyzer
Temperature between 10and 45°C (50-113F) at max 80% relative humidity. If
the temperature could fall below 5°C, the analyzer should be installed in a heated
cabinet.
below the analyzer for the reagent bottles.
Installation
24