3.4.7 AUTO CALIBRATION

The analyzer can be set up to perform an automatic 1-Pt or 2-Pt calibration during a
user-defined period.

3.4.8 CALIBRATION FAILURE

A Calibration Fail (CAL FAIL) condition will occur after a 2-Pt calibration if the electrode
response does not meet requirements. This happens when the electrodes "strength"
("slope") from a 2-Pt calibration is below 83 or greater than 110. This could be caused
by a number of factors (See next section - Troubleshooting).
NOTE: During single and two point calibrations sample needs to be delivered to analyzer
at flow rate of 150-400ml/min. If not, calibration will result in CF.