Vega VEGAWAVE 63 Operating Instructions Manual page 13

Vibrating level switch with tube extension for powders
Table of Contents

Advertisement

VEGAWAVE 63 • Transistor (NPN/PNP)
The tuning fork must be mounted in a way that takes the arrangement
of the filling and emptying apertures into account.
To compensate measurement errors caused by the material cone in
cylindrical vessels, the sensor must be mounted at a distance of d/6
from the vessel wall.
d
6
d
Fig. 4: Filling and emptying centred
4 Mounting
d
6
d
13