Siemens SITRANS F Instruction Manual page 140

Ultrasonic flowmeters
Hide thumbs Also See for SITRANS F:
Table of Contents

Advertisement

Section 2
4. Access the Installation Menu. Select [Meter Type] [Single, Dual or Quad Channel] depending on flow-
meter type. Select the meter channel (1,2,3, or 4) depending on which measurement channel you
intend to test. Select [Clamp-On] and then [Channel Setup].
5. Access the [Channel/Path Setup] menu. Move the highlight to [Create/Name Site]. Create a new Site
Setup (e.g., TEST1). You can now enter data without altering an existing Site Setup.
6. Select the [Pipe Data] Menu. Referencing either the English or metric pipe simulator chart below,
enter the pipe data corresponding to the transducer size under test.
1012TB-1 & 2 Universal Transducer Test Block Chart (English)
Part
Number
1012TB-1
1012TB-1
1012TB-2
1012TB-2
*System Defaults
Part
Number
1012TB1
1012TB1
1012TB2
1012TB2
*System Defaults
6. Access the [Pick/Install Xdcr] menu. Check the [Transducer Size] menu cell. If necessary, enter the
Xdcr Size option list and pick the transducer under test.
7. Check the [Xdcr Mount Mode] menu cell. Adjust to match the simulator chart above.
8. Move the highlight down to [Install Completed?]. Access the option list. Move the cursor to [Install].
Press <ENT> to start the transducer install routine.
NOTE: Since Sizes A and C transducers are installed in Reflect Mode, you will see a pop-up
window that prompts you to: [Use Actual Zero]. You can ignore this by pressing the
<Down Arrow>.
Upon the completion of the [Install] procedure, the flow computer should report a sonic velocity within
the range of approximately 1350m/s to 1700 m/s (depending on the transducer size under test). Next,
check [Diagnostic Data/Flow Data] menu. The [Valc %] item must be >35 for dependable operation. In
addition, note the reading, then compare it to the [Vfmax] item in the [Diagnostic Data/Site Setup
Data] menu. The value of the reading should be less that 2% of the published Vf max. Confirming these
values certifies that the entire system (computer, transducers, cables) is operating correctly. The inves-
tigation should proceed to a review of all site conditions to locate the operating problem.
Xdcr Pipe OD
Pipe
Size
(in.)
Mat'l
A
0.650
Steel*
B
1.150
Steel*
C
2.000
Steel*
D
3.500
Steel*
1012TB-1 & 2 Universal Transducer Test Block Chart (Metric)
Xdcr Pipe OD
Pipe
Size
(mm)
Mat'l
A
16.5
Steel*
B
29.2
Steel*
C
50.8
Steel*
D
88.9
Steel*
Wall
Mount
Thk (in.)
Mode
0.100*
Reflect
Nominal
0.100*
Direct
Minimum
0.100*
Reflect
Nominal
0.100*
Direct
Minimum
Wall
Mount
Thk (mm)
Mode
2.54*
Reflect
2.54*
Direct
2.54*
Reflect
2.54*
Direct
2-86
1010ENFM-3B
Spacing
Number
Offset
Index
7
4
11
Use Ltn
Spacing
Number
Offset
Index
Nominal
7
Minimum
4
Nominal
11
Minimum
Use Ltn