Safety Precautions Related to Installation Location
WARNING
• Some of the solvents used for measurement are flammable or toxic. Be sure to
ventilate the room properly.
Not doing so may result in poisoning or fire.
• Organic solvents are sometimes used for measurement. Therefore, do NOT use
open flames near the product and do NOT install the product in the same room as
equipment that generates sparks.
Doing so may result in fire.
As a safety measure, provide a fire extinguisher.
• A sink is required near the product
If solvent gets into your eyes or you come in contact with toxic solvent, you must wash away the solvent
immediately. Provide a sink as near the product as possible.
CAUTION
• Consider the weight of the entire measurement system when installing the product.
Install the product on a surface (e.g., a table) that is flat, stable, and easily capable of withstanding the
weight of the entire measurement system. Otherwise, the equipment may fall or topple over.
• Do NOT install the product in locations subject to large amounts of dirt, dust,
corrosive gases, or gases containing halides.
Otherwise, performance specifications may not be attained or the service life of the equipment may be
reduced.