Press and hold the
Step18 Press the
5.3. Weighing Range Function
The weighing range function can select "single range" or "dual range". Specify each
weighing interval (division) for the multi-interval instrument. Each weighing interval is
displayed according to a net value or gross value.
Caution
When single range is used, performing this function is not required.
Example 1
Specified parameters:
First range
Second range
Display
0kg to 50kg :
50kg to 100kg :
0kg
Example 2
Specified parameters:
Display
-40kg to 50kg :
50kg to 60kg :
-40kg
AD-4405A Weighing Indicator
key and press the
ON/OFF
No parameters are changed, CanCel is displayed and the calibration
mode is finished.
key to turn the display off.
ON/OFF
The gross display.
Range = 50.00kg, division 0.02kg
Range = 100.00kg (maximum capacity), division 0.1kg
The first range, division 0.02kg.
The second range, division 0.1kg.
Division = 0.02kg
The net display using a 40kg tare value.
The same parameters as example 1.
The first range, division 0.02kg.
The second range, division 0.1kg.
Division = 0.02kg
0kg
Tare value
40kg
key
ESC
Division = 0.1kg
50kg
Page 17
100kg Gross
Weighing value
Division=0.1kg
50kg 60kg Net
90kg 100kg Gross
Weighing value
5.3. Weighing Range Function