5.3. Weighing Range Function
The weighing range function can select "single range", "dual range" and "triple range".
Specify each weighing interval (division) for the multi-interval instrument. Each
weighing interval is displayed according to a net value or gross value.
Caution
Using a single range, it is not necessary to perform this function.
Perform the zero and span calibration after this function.
Example 1
Specified parameters:
First range
Second range
Third range
Display
0kg to 20kg
20kg to 50kg
50kg to 100kg :
Example 2
Specified parameters:
The same parameters as example 1.
Display
-40kg to -20kg :
-20kg to 20kg
20kg to 50kg
50kg to 60kg
5.3. Weighing Range Function
The gross display.
Range = 20.00kg, division 0.01kg
Range = 50.00kg, division 0.02kg
Range = 100.00kg (maximum capacity), division 0.1kg
:
The first range, division 0.01kg.
:
The second range, division 0.02kg.
The third range, division 0.1kg.
The net display using a 40kg tare value.
The second range, division 0.02kg.
:
The first range, division 0.01kg.
:
The second range, division 0.02kg.
:
The third range, division 0.1kg.