rmP.r (Ramp)
To ramp linearly at a set rate to a specified temperature:
1. Press the PAGE button until you reach the program list (ProG LiSt).
2. Press the SCROLL button until display reads,"tYPE."
3. Press the UP or DOWN button until display reads, "rmP.r."
Steps 4 and 5 are used in the 4 program model only. If you are using an 8 segment program, skip to
step 6.
4. Press the SCROLL button until display reads "Hb."
5. Press the UP or DOWN button to toggle between "bAnd, Hi, Lo and OFF."
6. Press the SCROLL button until display reads, "tGt."
7. Press the UP or DOWN button to set a target setpoint.
8. Press the SCROLL button until display reads,"rAtE."
9. Press the UP or DOWN button to select a value in ramp units (seconds, minutes or hours; set in
the "rmP.U" parameter).
rmP.t
To ramp to a specified temperature at a set time:
1. Press the PAGE button until you reach the program list (ProG LiSt).
2. Press the SCROLL button until display reads, "tYPE."
3. Press the UP or DOWN button until display reads, "rmP.t."
4. Press the SCROLL button until display reads, "tGt."
5. Press the UP or DOWN button to set a target setpoint.
6. Press the SCROLL button until display reads, "dur."
7. Press the UP or DOWN button to select a time in ramp units (seconds, minutes or hours; set in
the "rmP.U"parameter.
dwEll
To maintain a constant temperature for a specified time:
1. Press the PAGE button until you reach the program list (ProG LiSt).
2. Press the SCROLL button until display reads, "tYPE."
3. Press the UP or DOWN button until display reads, "dwEll."
4. Press the SCROLL button until display reads, "dur."
5. Press the UP or DOWN button to select a time in dwell units (seconds, minutes or hours; set in
the "dwL.U" parameter).
9
8 Segment & 4x16 Segment Programmable Models w/OTP
Programming the Controller