Combination Of Math Operations - Agilent Technologies U3402A User's And Service Manual

5 1/2 digit dual display multimeter
Hide thumbs Also See for U3402A:
Table of Contents

Advertisement

U3402-90001.book Page 48 Friday, July 24, 2009 4:04 PM
2
Operations and Features

Combination of Math Operations

N O TE
48
Downloaded from
Elcodis.com
electronic components distributor
The math operation can be operated for primary display only.
The Agilent U3402A multimeter allows you to use multiple math operation
such as dBm, MinMax, Rel, Hold, and Comp simultaneously.
Example:
Set upper and lower limit for Comp operation using multiple math operation
as below:
1 Press
to enable dBm operation.
2 Press
to take dBm readings as Rel # for new measurement.
3 Press
to record the Max value as the new Rel # (upper limit).
4 Press
to record the Min value as the new Rel # (lower limit).
5 Press
to enable the Comp operation.
The step- by- step sequence and readings of the combined math operations are
shown in
Table 2- 6
when all math operations are used sequentially. See
Figure 2-
19.