Minmax; Figure 2-15 Typical Max Operation Display - Agilent Technologies U3402A User's And Service Manual

5 1/2 digit dual display multimeter
Hide thumbs Also See for U3402A:
Table of Contents

Advertisement

U3402-90001.book Page 44 Friday, July 24, 2009 4:04 PM
2
Operations and Features
N O TE
44
Downloaded from
Elcodis.com
electronic components distributor
Procedure
1 Press
to enable Rel operation.
2 Press
3 Use
,
,
4 Press
to set the Rel # and read the display.
5 Press
to disable the Rel operation.
• In resistance measurement mode, the multimeter will read a non-zero value even when
the two test leads are in direct contact, because of the resistance of these leads. Use
the Rel operation to zero-adjust the display.
• In DC voltage measurement mode, the thermal effect will influence the accuracy. Short
the test leads and press
display.

MinMax

The MinMax (Minimum/Maximum) operation stores the minimum and
maximum values of reading during a series of measurements.
When enabled, the MinMax operation turns on the MINMAX annunciator and
begins accumulating various statistics of the readings being displayed.

Figure 2-15 Typical Max operation display

to toggle in (and out) the Rel #.
, and
to modify the Rel #.
once the displayed value is stable to zero-adjust the
U3402A User's and Service Guide