To Calculate Total Measurement Error - Agilent Technologies U3402A User's And Service Manual

5 1/2 digit dual display multimeter
Hide thumbs Also See for U3402A:
Table of Contents

Advertisement

U3402-90001.book Page 107 Friday, July 24, 2009 4:04 PM

To Calculate Total Measurement Error

U3402A User's and Service Guide
Downloaded from
Elcodis.com
electronic components distributor
The multimeter's accuracy specifications are expressed in the form:
(% of reading + count)
In addition to the reading error and count error, you may need to add
additional errors for certain operating conditions. Check the list below to make
sure you include all measurement errors for a given function. Also, make sure
you apply the conditions as described in the footnotes on the specification
pages.
• If you are operating the multimeter outside the temperature range
specified, apply an additional temperature coefficient error.
• For AC voltage and AC current measurements, you may need to apply
an additional low frequency error or crest factor error.
Specifications and Characteristics
6
107

Advertisement

Table of Contents
loading

Table of Contents