4.4 Cable Check

If you should have the impression that a cable is defective during treatment, you can check it using the in-
strument.
How to Check a Cable
(1)
Select a therapy and set the intensity to 100% using the Data Selector <2>.
(2)
Plug the ends of the cable to be tested into the Ports <3>.
(3)
Move the cable and pay attention to the Control Light <6> under the Data Selector <2>:
The Control Light <6> is permanently on: The cable is defective.
The Control Light <6> is off: The cable is OK.
The Control Light <6> flashes when moving the cable: The cable is defective.
(4)
Replace the defective cable immediately.

4.5 Handheld Applicator

CAUTION
Replace the treatment heads immediately if damaged. Any damage to the oscillator mem-
brane might lead to harmless (but unpleasant) sensations on the skin.
The handheld applicator consists of the following parts:
Hand Applicator (Exploded View)
The oscillator head is screwed into the handle. To remove the head from the handle, you must rotate it to
12
(1) Applicator Handhold (Ref.-No. 00379)
(2) Cover
(3) Oscillator Plate
(4) Membrane
(5) Clamp Ring
Important
(2) - (5) make up the oscillator head in
the respective size.