PHYSIOMED ELEKTROMEDIZIN AG

Appendix A Reference Information

A.1 Service, Repairs, Maintenance
NOTICE
The oscillator head must only be disinfected using an agent, but not be sterilized using heat.
All parts of the oscillator head must be absolutely dry before using them next time.
The manufacturer guarantees the safety of the instrument only in its original state. The instrument must
be operated in accordance with the Operating Instructions.
Repairs to the instrument may only be performed by Chudo or HIVAMAT
by PHYSIOMED ELEKTROMEDIZIN AG. Any repairs performed by an authorisedauthorized service center
must be accompanied by written certification, describing the nature and extent of the repairs undertaken,
as applicable with details regarding changes to nominal operating values or the operational range. The
certification must also contain the date performed, the name of the repair company and the signature of
the service person. When defective, components affecting the safe operation of the instrument must be
replaced by manufacturer's original parts. Upon request, wiring diagrams, parts lists and service instruc-
tions can be made available to qualified technical personnel employed by the customer.
We recommend having the instrument, including all accessories, serviced at regular intervals.
A.2 Cleaning and Disinfection
Clean the device and the accessories regularly with an aldehyde-based disinfectant. Switch off the instru-
ment by any means before cleaning.
Use a soft sponge cloth for cleaning. Ensure that the back cover is secured tightly and use caution so that
no liquid substances penetrate into the instrument.
For the hand applicator, use a new or disinfected oscillator head for each patient. Clean, disinfect or steril-
ize the hand applicator regularly. The clamping ring, oscillator plate and handle are resistant against disin-
fecting agents and can be sterilized up to 275°F (135°C).
A.3 Limited Warranty
Three-year limited warranty covers parts and labour on HIVAMAT
A.4 Disposal
After the service life of the instrument, dispose of it in conformance with the applicable regulations for en-
vironment protection.
Handle, clean, and dispose of components and accessories that have come in contact with bodily fluids ac-
20
Reference Information
200 service centers authorized
200 Personal.
HIVAMAT
200 Personal