Oven Power Cord And Fuses - PerkinElmer 200 Series User Manual

Table of Contents

Advertisement

2-4 Installing the Oven

Oven Power Cord and Fuses

The AC input power module is located on the left side of the back panel
and contains the on/off switch and system fuses.
Figure 2-1. Rear panel of the oven.
The unit is shipped with two 1.0 amp fuses installed for 230-Volt operation.
Prior to operation at 115 VAC, replace the fuses as described below.
To replace a fuse, proceed as follows (see Figure 2-1):
1. Turn off the AC power to the oven by pressing 0 on the On/Off switch (on
the rear of the oven). Then unplug the AC line cord from the AC outlet.
2. To remove a fuse, use a screwdriver to pry open the fuse holder. Install
2.0-amp fuses (P/N 0999-1637) in the fuse holder (for 115-Volt operation)
and reinsert the fuse holder in the power entry module.
3. Make certain that the Voltage Selector switch is in the position required.
(Here??)