Table 1-2 – Series 200 Peltier Column Oven, Oven/Selector Specifications
Specification
Column Capacity
Operating Range
Recommended Flow Range
Temperature Accuracy
Temperature Stability
Temperature Repeatability
Heating/Cooling Rates
Temperature Ramp
Safety Cutoff
Stabilizing Time
Power Requirements:
Power Consumption
Line Voltage Operating Range
Frequency
Line Cord
Fuse Rating Characteristics:
120 VAC (P/N 0998-1637)
240 VAC (P/N 0999-1656)
Operating Conditions:
Ambient Temperature
Ambient Relative Humidity (RH)
Altitude
Storage Conditions:
Ambient Temperature
Ambient Relative Humidity
Altitude
Dimensions
Weight
*With ambient temperatures of between 15°C and 20°C.
**With ambient temperatures of between 20°C and 25°C.
***With ambient temperatures of between 25°C and 30°C.
Description
Up to three 3/8-in. × 30-cm columns or three ¼-in ×
30-cm columns
5°C to 90°C (39°F to 194°F)
Up to 5 mL/minute
± 1°C between 5°C and 90°C*
± 1°C between 10°C and 90°C**
± 2°C between 15°C and 90°C***
± 0.2°C (± .36°F)
± 1°C (1.8°F)
30 minutes to 10°C (50°F) at 2 mL/min* or
30 minutes to 60°C (140°F) at 2 mL/min*
Up to 5 °C/min (9 °F/min) cooling and heating*
Leakage sensor
30 minutes after making setting
120 Watts
115 or 230 Vac, 50/60 Hz ±10% (Line voltage is set at
the factory to 230 Vac.)
50/60 Hz
See "Installing the AC Line Cord" in Chapter 2
2.0 Amp, Type T, 250 Volt
1.0 Amp, Type T, 250 Volt
10 to 35°C (50 to 95°F)
20 to 80% RH (without condensation)
0 to 2000 m
-20 to +60°C (-4 to+140°F)
20 to 80% RH (without condensation)
0 to 12,000 m
Height: 16.0 cm (6.3 in.)
Width: 34.0 cm (13.4 in.)
Depth: 48.0 cm (18.9 in.)
9 Kg (20 lb)