Installing the Oven

Unpack Your Series 200 Peltier Column Oven

Carefully unpack the Series 200 Peltier Column Oven. Then check for
obvious signs of damage that may have occurred during shipment. Immediately
report any damage or missing items to the shipping carrier and PerkinElmer.
A Start-Up Kit is supplied with the oven. Use Table 2-1 to check the Start-
Up Kit parts.
Table 2-1. Start-Up Kit for the Series 200 Peltier Column Oven
AC Line Cord*
Fingertight Nuts
Inlet Union
2-Amp Fuses
1-Amp Fuses
Instruction Manual
* Customers located outside the U.S. must order a line cord that is appropriate for their location

Electrical Requirements and Settings

The Series 200 Peltier Column Oven can operate at 50 or 60 Hz and
requires a grounded outlet that provides a nominal 115 or 230 Vac source.
If the electrical supply voltage in your laboratory produces large AC line
voltage fluctuations, install a voltage regulator between the AC line voltage
outlet and your LC system. Instrument grounding is required and
accomplished by a third wire in the AC line voltage outlet box. If your
electrical outlet does not contain a ground, contact your local electric
company.
Items Included

Unpack Your Series 200 Peltier Column Oven

P/N
Quantity
0998-8986
0992-0513
0990-3289
0999-1637
0999-1656
0993-6398
2-3
1
2
1
2
2
1