Pressure Buildup

Over time, you may notice a gradual increase in the system operating
pressure. If you observe pressure readings greater than 3.45 MPa (500 psi)
above the normal operating pressure of your analysis, check the following
items:
If you are injecting "clean" samples and there is no other apparent cause
for the pressure buildup, remove, disassemble, and clean the injector.
If you are injecting "dirty" samples, the injector may be clogged or the
column packing material may have retained contaminated particulates.
Remove and clean the injector, and replace the column.
The column end fitting or column frit may be plugged. Replace the end
fitting or frit to relieve excess pressure. (Refer to the instructions
supplied with the column.)
In-line filters may be plugged. Replace the filter element.
With the pump connected directly to the detector input, if you still
experience excessive pressure, the problem could be a plugged flowcell.
You can flush, reverse-flush, or rebuild the flowcell.