10. Specifications
10.3 Model CPH6200 hand-held pressure indicator (complete measuring chain)
Model CPH6200 hand-held pressure indicator (complete measuring chain)
Accuracy of the
EN
measuring chain
Mean temperature
coefficient
Compensated range
1) It is defined by the total measurement uncertainty, which is expressed with the coverage factor (k = 2) and includes the following
factors: the intrinsic performance of the instrument, the measurement uncertainty of the reference instrument, long-term stability,
influence of ambient conditions, drift and temperature effects over the compensated range during a periodic zero point correction.
2) Reference conditions: 15 ... 25 °C [59 ... 77 °F]
10.4 Approvals
CE conformity
Model CPH6200
Model CPT6200
For further approvals, see website
10.5 Certificates
Certificate
Calibration for
model CPT6200
Recommended
calibration interval
For certificates, see website
For further specifications, see WIKA data sheet CT 11.01 and the order documentation.
42
0.2 % FS
■
1)
0.1 % FS at reference conditions
■
< 100 mbar [< 1.45 psi])
≤ 0.2 % of span/10 K (outside the reference conditions)
0 ... 80 °C [0 ... 176 °F]
EMC directive
■
EN 61326 emission (group 1, class B) and immunity (portable
equipment)
RoHS directive
■
EMC directive
■
EN 61326 emission (group 1, class B) and immunity (portable
measuring equipment)
Pressure equipment directive
■
Module A, internal production control
RoHS directive
■
3.1 inspection certificate per EN 10204 (factory calibration)
■
DAkkS calibration certificate (traceable and accredited in
■
accordance with ISO/IEC 17025)
1 year (dependent on conditions of use)
2)
(not for pressure ranges
WIKA operating instructions, model CPH6200
2)
Need help?
Do you have a question about the CPH6200 and is the answer not in the manual?