Portable digital pressure gauge

Advertisement

Quick Links

Operating Instructions
Portable Digital Pressure Gauge
CPG2300
Portable Digital Pressure Gauge CPG2300 Series
PN 0017441001M • 10/2014

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the CPG2300 Series and is the answer not in the manual?

Questions and answers

Summary of Contents for WIKA CPG2300 Series

  • Page 1 Operating Instructions Portable Digital Pressure Gauge CPG2300 Portable Digital Pressure Gauge CPG2300 Series PN 0017441001M • 10/2014...
  • Page 2 This Caution symbol indicates danger for the system and material if the respective safety precautions are not taken. Caution This Notice symbol does not indicate safety notices but information for a better understanding of the facts. Operating Instructions CPG2300 Series...
  • Page 3: Table Of Contents

    5.1 Unpacking the Instrument 5.2 Dimensions 5.3 Pressure Connections 5.4 Electrical Connections 5.5 Power On 6. Local Operation 6.1 Keypad 6.2 Standard Display 6.3 Menu Screens 6.3.1 Main Menu Screen 6.3.2 Sensor Setup Menu 6.3.3 Password Entry Screen Operating Instructions CPG2300 Series...
  • Page 4 8.1 Current Input 8.1.1 Menu Screens 8.1.2 Operational Notes 8.1.3 Connection Configurations 8.1.4 Application Diagrams 9. Calibration 9.1 Environment 9.2 Equipment 9.3 Pressure Standard 9.4 Calibration Medium 9.5 Calibration Process 9.5.1 Zero Adjustment 9.5.2 Span Adjustment Operating Instructions CPG2300 Series...
  • Page 5: General Information

    Mensor has made a concerted effort to provide complete and current information for the proper use of the equipment. If there are questions regarding this manual or the proper use of the equipment, contact either Mensor or WIKA: Mensor WIKA Alexander Wiegand SE & Co. KG 201 Barnes Drive Alexander-Wiegand-Straße 30...
  • Page 6: Software License Agreement

    1.4.3 Certifications and Accreditations Mensor is registered to ISO 9001:2008. The calibration program at Mensor is accredited by A2LA, as complying with both the ISO/IEC 17025:2005 and the ANSI/NCSL Z540-1- 1994 standards. Operating Instructions CPG2300 Series...
  • Page 7: Safety Notices

    CAUTION: ESD PROTECTION REQUIRED. The proper use of ground- ed work surfaces and personal wrist straps are required when coming into contact with exposed circuits (printed circuit boards) to prevent static discharge to sensitive electronic components. Additional notices are found throughout this manual. Operating Instructions CPG2300 Series...
  • Page 8 Portable Digital Pressure Gauge CPG2300 Series Notes Operating Instructions CPG2300 Series...
  • Page 9: Product Description

    There is a battery level indicator on the display that indicates the battery charge level. The battery level indicator will blink when there is approximate- ly one hour of battery left. The unit will automatically shut off when the battery reaches a predetermined level. Operating Instructions CPG2300 Series...
  • Page 10: Serial In/Out Communications

    CPG2300 or to read the output of its sensors. Wiring requirements are provided in the Installation section and a list of valid commands and responses (Table 7.2.3) is provided in the Remote Operation section. Operating Instructions CPG2300 Series...
  • Page 11: Specifications

    Ranges ≥ 5 psi: Media compatible with aluminum, 316 SS, Buna N, Viton, silicone grease and RTV. Cannot guarantee accuracy on media other than gases. Not designed for oxygen service. Reference port: Clean, dry, non-corrosive gases. Operating Instructions CPG2300 Series...
  • Page 12 @0C and 60F, inH2O @4C, 20C, bar, mbar, mmHg @ 0C, cmHg @0C, hPa, kPa, Pa, MPa, kg/sq cm, cmH2O @4C, cmH2O @20C, mSW @0C. Display Monochrome 128 x64 LCD, with white LED back light. Compliant to EN50081 and EN50082. Operating Instructions CPG2300 Series...
  • Page 13: Installation

    Upon receipt, please examine the transducer for shipping damage. Report any apparent damage to the carrier immediately. In addition to this manual you should have: • CPG2300 Digital Pressure Gauge; • Power adapter (charger) • Any accessories ordered; • An envelope containing the Calibration Certificate. Operating Instructions CPG2300 Series...
  • Page 14: Dimensions

    Portable Digital Pressure Gauge CPG2300 Series 5.2 Dimensions Figure 5.2 - Case Outline/Dimensions Operating Instructions CPG2300 Series...
  • Page 15: Pressure Connections

    CAUTION: It is recommended that a backup wrench be used when installing fittings or connections on the CPG2300. Caution Sense + INPUT JACK (red) Sense - INPUT JACK (black) (for Input Current Option) Figure 5.3 - Pressure and Electrical Connections Operating Instructions CPG2300 Series...
  • Page 16: Electrical Connections

    5.5 Power On After the pressure connections are secure, press the power switch ( ). The CPG2300 will run through a brief initialization sequence and then will display the sensor channel(s) that was active when turned off last. Operating Instructions CPG2300 Series...
  • Page 17: Local Operation

    The “Main Menu” Key Power Switch The M key beings up the main This key is the power on/off menu. switch. Figure 6.1 - Keypad and Display Features Operating Instructions CPG2300 Series...
  • Page 18: Standard Display

    Sensor A: Displays only the A Sensor. Sensor B: Displays only the B Sensor. Sensor A&B: Displays both the A and the B Sensor. Sensor A setup: Displays the Sensor A setup screen. Sensor B setup: Displays the Sensor B setup screen. Operating Instructions CPG2300 Series...
  • Page 19: Sensor Setup Menu

    Peak readings are displayed only when a single channel is selected. Peak Reset: Resets the peak values to the active measured pressure. A Setup Peak Off Peak Rst Null Off Units Operating Instructions CPG2300 Series...
  • Page 20 Sensor A Calibration A Setup Password Entry Peak Off Peak Rst Null Off Enter to Continue Units Units: Displays the first of the three units screens. InHg 0c InHg 60f InH20 4c InH20 20c mBAR more Operating Instructions CPG2300 Series...
  • Page 21: Password Entry Screen

    Adjust the digits to the true pressure applied to the MA Calibration active sensor and press the ♦ (enter) key to store the Cal Zero Cal Span value. For more details, see the Calibration section. Change PW Reset Cal Operating Instructions CPG2300 Series...
  • Page 22 Other Keys to Quit If the ♦ (enter) key is pressed, the calibration is reset to the factory defaults and the main operation screen is displayed. If any other key is pressed, the main menu screen is displayed. Operating Instructions CPG2300 Series...
  • Page 23: Units Screens

    To apply the “user” conversion factor, the units “User” must be selected for the active sensor. For the remote programmable pressure units list and information, see “User Programmable Pressure Units List” in the Remote Operation section of this manual. Operating Instructions CPG2300 Series...
  • Page 24: Menu Tree

    Portable Digital Pressure Gauge CPG2300 Series 6.4 Menu Tree CPG2300 Figure 6.4 - CPG2300 Menu Tree Operating Instructions CPG2300 Series...
  • Page 25: Remote Operation

    (A or A&B) the Channel A is active. In the case of a single chan- nel unit, Channel A is always active. It is good practice to query for the active channel before sending commands that change a sensors operating parameters. See Table 7.2.3 for commands and queries. Operating Instructions CPG2300 Series...
  • Page 26: Password Protection

    Address (single character 0 through 9 or A through Z; case insensitive) Global address character Query operator <sp> Space ( ) Password (insert password in place of PW; case insensitive) One character place holder for digit or decimal point n...n Variable length field Ready response Operating Instructions CPG2300 Series...
  • Page 27: Commands And Queries

    +14.456, +14.300<cr><lf>. #XA<sp>n<cr><lf> Sets address of CPG2300 to 0-9 or A-Z. #XCH?<cr><lf> X<sp>n<cr><lf> Returns active channel “A” or “B”. #XCH<sp>n<cr> Sets active sensor channel where <lf> n is A or B. #XCPW<sp>nn Enter customer password. nn<cr><lf> Operating Instructions CPG2300 Series...
  • Page 28 Available only with the Current Measurement option; Returns current measurement in mil- liamps. #XID?<cr><lf> X<sp>ID<sp>nn Returns the CPG2300 ID string. MENSOR, <sp>nnnn 64nn,<sp>nnnnnnnn, <sp>Vn.nn<cr><lf> NOS? #XNOS?<cr><lf> X<sp>NOS<sp>n Returns number of sensors <cr><lf> installed in the CPG2300. Operating Instructions CPG2300 Series...
  • Page 29 Sets the units code for the active sensor. USR? #XUSR?<cr><lf> X<sp>USR<sp>nn Returns the user unit multiplier nnnnnnnennnn for the active sensor. <cr><lf> #XUSR<sp>n Sets user selected units for the <cr><lf> active sensor. Ranges from -1.00000e+6 to -1.00000e-6 and from +1.00000e-6 to +1.00000e+6. Operating Instructions CPG2300 Series...
  • Page 30: User Programmable Pressure Units List

    #*UT 2 34<cr><lf> - the third unit displayed would be hPa #*UT 3 0<cr><lf> - terminates the displayed units to this point #*UT 0 0<cr><lf> - allows all factory units to be displayed and used The “SAVE” command must be sent to make changes permanent. Notice Operating Instructions CPG2300 Series...
  • Page 31: Pressure Unit Codes And Conversions

    4C 27.680672 inH2O 20C 27.729767 0.06894757 mbar 68.94757 cmH2O 4C 70.3089 mmHg 0C 51.71508 cmHg 0C 5.171508 6.894757 6894.757 kg/sq cm 0.07030697 mSW @0C (@3.5% salinity) 0.6838528 68.94757 0.006894757 cmH2O 20C 70.4336 User User Settable Operating Instructions CPG2300 Series...
  • Page 32 Portable Digital Pressure Gauge CPG2300 Series Notes Operating Instructions CPG2300 Series...
  • Page 33: Options

    Input Setup: The items in this menu perform the following functions when selected: mA Input Setup Display mA Off Source V Off Calibrate mA Display mA Off (On): Enables/disables current measurement by toggling the enter key. mA Input Setup Display mA On Source V Off Calibrate mA Operating Instructions CPG2300 Series...
  • Page 34 Calibration Screen: The mA calibration screen allows zero and span to be adjust- ed, the calibration to be reset to the factory defaults or the calibration password to be changed. MA Calibration Cal Zero Cal Span Change PW Reset Cal Operating Instructions CPG2300 Series...
  • Page 35: Operational Notes

    This is accomplished by issuing the “I?” command (see Remote Operation section). The peak pressure indication function can not be enabled when the current input is enabled. The null feature will work when the current input is enabled. Operating Instructions CPG2300 Series...
  • Page 36: Connection Configurations

    “Sense –” connection each through 1 Meg ohm resistors. The RS-232 digital ground is connected directly to the “Sense –”. The connections should be isolated from the current loop connections. Figure 8.1.3 - Electrical Diagram (Current Input option) Operating Instructions CPG2300 Series...
  • Page 37: Application Diagrams

    Portable Digital Pressure Gauge CPG2300 Series 8.1.4 Application Diagrams Figure 8.1.4a - Connections in a powered loop Figure 8.1.4b - Connections using CPG2300 loop power Operating Instructions CPG2300 Series...
  • Page 38 Portable Digital Pressure Gauge CPG2300 Series Notes Operating Instructions CPG2300 Series...
  • Page 39: Calibration

    Figure 9.2. These items are: 6. A vacuum pump capable of pumping down to 300 millitorr. 7. A vacuum standard to measure pressures down to 300 millitorr. 8. One additional valve to isolate the vacuum line. Operating Instructions CPG2300 Series...
  • Page 40 Portable Digital Pressure Gauge CPG2300 Series CPG2300 PRESSURE GAUGE CPG2300 PRESSURE GAUGE Figure 9.2 - Calibration Setup Operating Instructions CPG2300 Series...
  • Page 41: Pressure Standard

    Press the ♦ (enter) key to store the zero offset. For a bi-directional sensor: Apply a negative pressure source equal to R-. Operating Instructions CPG2300 Series...
  • Page 42: Span Adjustment

    50% of the span or span adjustments will not be accepted. Adjust the measured pressure on the span calibration screen to match the true applied pressure. Press the ♦ (enter) key to store the span offset data. Operating Instructions CPG2300 Series...
  • Page 43 Portable Digital Pressure Gauge CPG2300 Series Notes Operating Instructions CPG2300 Series...
  • Page 44 512.396.4200 fax: 512.396.1820 web site: www.mensor.com email: sales@mensor.com Operating Instructions CPG2300 Series PN 0017441001M • 10/2014 WIKA Alexander Wiegand SE & Co. KG Alexander-Wiegand-Straße 30 D-63911 Klingenberg / Germany tel: (+49) 93 72/132-9986 web site: www.wika.de fax: (+49) 93 72/132-8767...