Operating Instructions Portable Digital Pressure Gauge CPG2300 Portable Digital Pressure Gauge CPG2300 Series PN 0017441001M • 10/2014...
Page 2
This Caution symbol indicates danger for the system and material if the respective safety precautions are not taken. Caution This Notice symbol does not indicate safety notices but information for a better understanding of the facts. Operating Instructions CPG2300 Series...
5.1 Unpacking the Instrument 5.2 Dimensions 5.3 Pressure Connections 5.4 Electrical Connections 5.5 Power On 6. Local Operation 6.1 Keypad 6.2 Standard Display 6.3 Menu Screens 6.3.1 Main Menu Screen 6.3.2 Sensor Setup Menu 6.3.3 Password Entry Screen Operating Instructions CPG2300 Series...
Page 4
8.1 Current Input 8.1.1 Menu Screens 8.1.2 Operational Notes 8.1.3 Connection Configurations 8.1.4 Application Diagrams 9. Calibration 9.1 Environment 9.2 Equipment 9.3 Pressure Standard 9.4 Calibration Medium 9.5 Calibration Process 9.5.1 Zero Adjustment 9.5.2 Span Adjustment Operating Instructions CPG2300 Series...
Mensor has made a concerted effort to provide complete and current information for the proper use of the equipment. If there are questions regarding this manual or the proper use of the equipment, contact either Mensor or WIKA: Mensor WIKA Alexander Wiegand SE & Co. KG 201 Barnes Drive Alexander-Wiegand-Straße 30...
1.4.3 Certifications and Accreditations Mensor is registered to ISO 9001:2008. The calibration program at Mensor is accredited by A2LA, as complying with both the ISO/IEC 17025:2005 and the ANSI/NCSL Z540-1- 1994 standards. Operating Instructions CPG2300 Series...
CAUTION: ESD PROTECTION REQUIRED. The proper use of ground- ed work surfaces and personal wrist straps are required when coming into contact with exposed circuits (printed circuit boards) to prevent static discharge to sensitive electronic components. Additional notices are found throughout this manual. Operating Instructions CPG2300 Series...
Page 8
Portable Digital Pressure Gauge CPG2300 Series Notes Operating Instructions CPG2300 Series...
There is a battery level indicator on the display that indicates the battery charge level. The battery level indicator will blink when there is approximate- ly one hour of battery left. The unit will automatically shut off when the battery reaches a predetermined level. Operating Instructions CPG2300 Series...
CPG2300 or to read the output of its sensors. Wiring requirements are provided in the Installation section and a list of valid commands and responses (Table 7.2.3) is provided in the Remote Operation section. Operating Instructions CPG2300 Series...
Ranges ≥ 5 psi: Media compatible with aluminum, 316 SS, Buna N, Viton, silicone grease and RTV. Cannot guarantee accuracy on media other than gases. Not designed for oxygen service. Reference port: Clean, dry, non-corrosive gases. Operating Instructions CPG2300 Series...
Page 12
@0C and 60F, inH2O @4C, 20C, bar, mbar, mmHg @ 0C, cmHg @0C, hPa, kPa, Pa, MPa, kg/sq cm, cmH2O @4C, cmH2O @20C, mSW @0C. Display Monochrome 128 x64 LCD, with white LED back light. Compliant to EN50081 and EN50082. Operating Instructions CPG2300 Series...
Upon receipt, please examine the transducer for shipping damage. Report any apparent damage to the carrier immediately. In addition to this manual you should have: • CPG2300 Digital Pressure Gauge; • Power adapter (charger) • Any accessories ordered; • An envelope containing the Calibration Certificate. Operating Instructions CPG2300 Series...
CAUTION: It is recommended that a backup wrench be used when installing fittings or connections on the CPG2300. Caution Sense + INPUT JACK (red) Sense - INPUT JACK (black) (for Input Current Option) Figure 5.3 - Pressure and Electrical Connections Operating Instructions CPG2300 Series...
5.5 Power On After the pressure connections are secure, press the power switch ( ). The CPG2300 will run through a brief initialization sequence and then will display the sensor channel(s) that was active when turned off last. Operating Instructions CPG2300 Series...
The “Main Menu” Key Power Switch The M key beings up the main This key is the power on/off menu. switch. Figure 6.1 - Keypad and Display Features Operating Instructions CPG2300 Series...
Sensor A: Displays only the A Sensor. Sensor B: Displays only the B Sensor. Sensor A&B: Displays both the A and the B Sensor. Sensor A setup: Displays the Sensor A setup screen. Sensor B setup: Displays the Sensor B setup screen. Operating Instructions CPG2300 Series...
Peak readings are displayed only when a single channel is selected. Peak Reset: Resets the peak values to the active measured pressure. A Setup Peak Off Peak Rst Null Off Units Operating Instructions CPG2300 Series...
Page 20
Sensor A Calibration A Setup Password Entry Peak Off Peak Rst Null Off Enter to Continue Units Units: Displays the first of the three units screens. InHg 0c InHg 60f InH20 4c InH20 20c mBAR more Operating Instructions CPG2300 Series...
Adjust the digits to the true pressure applied to the MA Calibration active sensor and press the ♦ (enter) key to store the Cal Zero Cal Span value. For more details, see the Calibration section. Change PW Reset Cal Operating Instructions CPG2300 Series...
Page 22
Other Keys to Quit If the ♦ (enter) key is pressed, the calibration is reset to the factory defaults and the main operation screen is displayed. If any other key is pressed, the main menu screen is displayed. Operating Instructions CPG2300 Series...
To apply the “user” conversion factor, the units “User” must be selected for the active sensor. For the remote programmable pressure units list and information, see “User Programmable Pressure Units List” in the Remote Operation section of this manual. Operating Instructions CPG2300 Series...
(A or A&B) the Channel A is active. In the case of a single chan- nel unit, Channel A is always active. It is good practice to query for the active channel before sending commands that change a sensors operating parameters. See Table 7.2.3 for commands and queries. Operating Instructions CPG2300 Series...
Address (single character 0 through 9 or A through Z; case insensitive) Global address character Query operator <sp> Space ( ) Password (insert password in place of PW; case insensitive) One character place holder for digit or decimal point n...n Variable length field Ready response Operating Instructions CPG2300 Series...
+14.456, +14.300<cr><lf>. #XA<sp>n<cr><lf> Sets address of CPG2300 to 0-9 or A-Z. #XCH?<cr><lf> X<sp>n<cr><lf> Returns active channel “A” or “B”. #XCH<sp>n<cr> Sets active sensor channel where <lf> n is A or B. #XCPW<sp>nn Enter customer password. nn<cr><lf> Operating Instructions CPG2300 Series...
Page 28
Available only with the Current Measurement option; Returns current measurement in mil- liamps. #XID?<cr><lf> X<sp>ID<sp>nn Returns the CPG2300 ID string. MENSOR, <sp>nnnn 64nn,<sp>nnnnnnnn, <sp>Vn.nn<cr><lf> NOS? #XNOS?<cr><lf> X<sp>NOS<sp>n Returns number of sensors <cr><lf> installed in the CPG2300. Operating Instructions CPG2300 Series...
Page 29
Sets the units code for the active sensor. USR? #XUSR?<cr><lf> X<sp>USR<sp>nn Returns the user unit multiplier nnnnnnnennnn for the active sensor. <cr><lf> #XUSR<sp>n Sets user selected units for the <cr><lf> active sensor. Ranges from -1.00000e+6 to -1.00000e-6 and from +1.00000e-6 to +1.00000e+6. Operating Instructions CPG2300 Series...
#*UT 2 34<cr><lf> - the third unit displayed would be hPa #*UT 3 0<cr><lf> - terminates the displayed units to this point #*UT 0 0<cr><lf> - allows all factory units to be displayed and used The “SAVE” command must be sent to make changes permanent. Notice Operating Instructions CPG2300 Series...
Input Setup: The items in this menu perform the following functions when selected: mA Input Setup Display mA Off Source V Off Calibrate mA Display mA Off (On): Enables/disables current measurement by toggling the enter key. mA Input Setup Display mA On Source V Off Calibrate mA Operating Instructions CPG2300 Series...
Page 34
Calibration Screen: The mA calibration screen allows zero and span to be adjust- ed, the calibration to be reset to the factory defaults or the calibration password to be changed. MA Calibration Cal Zero Cal Span Change PW Reset Cal Operating Instructions CPG2300 Series...
This is accomplished by issuing the “I?” command (see Remote Operation section). The peak pressure indication function can not be enabled when the current input is enabled. The null feature will work when the current input is enabled. Operating Instructions CPG2300 Series...
“Sense –” connection each through 1 Meg ohm resistors. The RS-232 digital ground is connected directly to the “Sense –”. The connections should be isolated from the current loop connections. Figure 8.1.3 - Electrical Diagram (Current Input option) Operating Instructions CPG2300 Series...
Portable Digital Pressure Gauge CPG2300 Series 8.1.4 Application Diagrams Figure 8.1.4a - Connections in a powered loop Figure 8.1.4b - Connections using CPG2300 loop power Operating Instructions CPG2300 Series...
Page 38
Portable Digital Pressure Gauge CPG2300 Series Notes Operating Instructions CPG2300 Series...
Figure 9.2. These items are: 6. A vacuum pump capable of pumping down to 300 millitorr. 7. A vacuum standard to measure pressures down to 300 millitorr. 8. One additional valve to isolate the vacuum line. Operating Instructions CPG2300 Series...
Press the ♦ (enter) key to store the zero offset. For a bi-directional sensor: Apply a negative pressure source equal to R-. Operating Instructions CPG2300 Series...
50% of the span or span adjustments will not be accepted. Adjust the measured pressure on the span calibration screen to match the true applied pressure. Press the ♦ (enter) key to store the span offset data. Operating Instructions CPG2300 Series...
Page 43
Portable Digital Pressure Gauge CPG2300 Series Notes Operating Instructions CPG2300 Series...
Page 44
512.396.4200 fax: 512.396.1820 web site: www.mensor.com email: sales@mensor.com Operating Instructions CPG2300 Series PN 0017441001M • 10/2014 WIKA Alexander Wiegand SE & Co. KG Alexander-Wiegand-Straße 30 D-63911 Klingenberg / Germany tel: (+49) 93 72/132-9986 web site: www.wika.de fax: (+49) 93 72/132-8767...