Switching off the EM
 Ensure the beam is off
 Turn off HT
 Close TEM Center
 Shut down computer
 Press POWER OFF (Right hand door). Lights go off under the buttons.
 Wait till microscope completely shut down.
 Switch off at the wall.
After a 'crash'
 If the HP PC is still on, switch off
 Switch off the AMT computer
 Switch on the JEM 1400 as below
Switch on the JEM 1400
 Press the
button. EM will start.
 Switch on the AMT computer and camera control
 Switch on the HP computer.
 Log on using the HP keyboard and JEOL MOUSE at rear of EM.
 Check for communications.
 TEM Center loads automatically.
 Check vacuum status to ensure the correct valves are open.
 Ensure the lenses are on
 Wait RT1 (PiG5) ~119µA