JEOL JFC-1100 User Instructions

Fine coat ion sputter

Advertisement

Quick Links

Jeol Fine Coat Ion Sputter JFC-1100
User instructions
Operation of this equipment is restricted to trained personnel familiar with the user
instructions, general safety precautions, and laboratory rules. Contact the laboratory
manager or your supervisor/group leader to get permission to use this equipment.
Chamber
Voltmeter
Ammeter
Updated | Oppdatert 10.07.2018

Advertisement

loading

Summary of Contents for JEOL JFC-1100

  • Page 1 Jeol Fine Coat Ion Sputter JFC-1100 User instructions Operation of this equipment is restricted to trained personnel familiar with the user instructions, general safety precautions, and laboratory rules. Contact the laboratory manager or your supervisor/group leader to get permission to use this equipment.
  • Page 2 Routine sample coating Use gloves to handle the samples in order to avoid contamination of your sample and of the chamber. 1. Connect the power cord to the electricity plug. 2. Open the chamber and place the sample(s) on the sample holder (Fig.1). Fig.
  • Page 4 Experimental Conditions For the preparation/coating of conodont specimens, the coating conditions are as follows:  Voltage: 1200 V (5–10 mA) and coating time of 4 min  Process is stopped and stubs are rotated by ~90°  Continue coating for 4 min In case of severe charging of conodont specimens, the following coating conditions have been followed: ...