5.4 Installation

Installation takes place under the direction and responsibility of a qualified technician of Pam
Mobility s.r.l.
ATTENTION: it is absolutely forbidden to assemble and install the stretcher without
the support of a qualified technician of Pam Mobility s.r.l. Similarly, it is absolutely
forbidden to disassemble the stretcher for subsequent reinstallation without the
support of a qualified technician of Pam Mobility s.r.l.
Check that the installation surface is sufficient considering the additional space necessary for

assembly.

Make sure that the space left next to the stretcher is sufficient for a person to pass by.
Make sure that the specific floor capacity is sufficient to support the weight of the stretcher with
the safe work load applied.
5.4.1 Preparing the installation area
The place of installation must have a rigid, horizontal, flat floor.

5.5 Checking the equipment

The packaging contains:
▪ PULSE stretcher (ordered model);
▪ additional accessories ordered;
▪ the user instruction manual.
5.6 Assembly
ATTENTION! The assembly area must be clean and clear; it must be at least 4x3 m
to allow assembly operations.
The place of assembly must have the following characteristics:
▪ flat, non-yielding floor;
▪ 400 LUX lighting.

5.7 Electrical connection

WARNING! The electric stretcher cannot be used in a potentially explosive or
flammable atmosphere (such as a hyperbaric chamber).
ATTENTION! Danger of electric shock. The cables must be positioned in such a
way that they cannot be crushed, trapped, pulled taut, walked on, bent, become
wet or obstructed with respect to the moving parts.
WARNING! The power cable must not cause obstruction to the operator.
WARNING! Check that the mains voltage and frequency correspond to that
indicated on the identification plate.
▪ Prepare a SCHUKO type socket;
▪ connect the plug to the power supply;
▪ the stretcher must be charged for at least 12 hours before use.