PamMobility SYNAGO Instructions For Use And Maintenance Manual

Bed with variable height

Advertisement

Quick Links

SYNAGO BED WITH VARIABLE
HEIGHT
727T0045 - 727T0045C

Advertisement

Table of Contents
loading

Summary of Contents for PamMobility SYNAGO

  • Page 1 SYNAGO BED WITH VARIABLE HEIGHT 727T0045 - 727T0045C...
  • Page 2 REVISION TABLE Revision Data Note 15/11/2016 First edition 02/07/2018 02/07/2018 Update general description, technical Update general description, data, sides, lower section of the legs technical data, sides, lower section of the legs Insertion variant 727T0045S 23/10/2019 14/04/2021 CE marking according to Regulation (EU) 2017/745 22/03/2023 Insertion variant 727T0045C...
  • Page 3: Table Of Contents

    Summary Summary ................3 GENERAL PROVISIONS ............7 Presentation of the handbook ..............7 Services with reduced costs ..............7 Conventions ..................7 GENERAL POWERS ............. 8 Constructor ..................8 Intended use ..................8 Essential performance of the product ............ 8 Environmental limitation of use ...............
  • Page 4 6.5 Adequacy check ................22 6.6 Assembly ..................23 6.7 Mounting Side Panels( cod . 7300007A) ..........23 6.8 Electrical connection ................24 6.9 Test funzionale ................... 24 FUNCTIONING ..............25 7.1 Warnings ..................25 7.2 Safe location ..................25 7.3 Emergency positions ................25 7.4 Move ....................
  • Page 5 9.2 Sanitization with products containing halogens ......52 9.3 Sanification with food additives ............52 9.4 Intervallinification ................53 9.5 Automatic sanitization ................. 53 9.6 Manual sanitization ................53 10. MAINTENANCE ............. 54 10.1 Periodic check ................... 54 10.2 Technical assistance ................54 10.3 Provision ..................
  • Page 6 ID08435 - Rev. 1.2 - 19/04/2023...
  • Page 7: General Provisions

    1. GENERAL PROVISIONS Presentation of the manual This manual is intended to provide the user with all the necessary information so that, in addition to proper use of the device, it is able to manage the same in the most autonomous and safe way possible.
  • Page 8: Constructor

    Via Verdi 39 - 42043 Gattatico (RE) - Italy Tel 0522 473859 - Fax 0522 1548244 E-mail: info@pammobility.comhttp: www.pammobility.com Intended use ▪ Device type: electric bed. ▪ The device is intended for use in the diagnosis, treatment and monitoring of an adult patient* under the strict supervision of medical personnel.
  • Page 9 requirements of this manual and contacting the assistance of Pam Mobility s.r.l. whenever a bed failure occurs. After 10 years of use it is recommended to replacethe interoletto. ID08435 - Rev. 1.2 - 19/04/2023...
  • Page 10: Identifying

    Identifying ATTENTION! It is forbidden to remove the label from the device for any reason. The article is identifiable by the plate on the base in which the following data are shown: A. Company logo; B. Device description; C. Article code; D.
  • Page 11: Safety

    SAFETY Safety rules 3.1.1 Definitions The following terms shall be used in this Safety Manual: Operator: the person in charge of installing, operating, regular performing maintenance, cleaning, repairing, transporting the device. Technic Pam mobility: qualified technician made available by Pam Mobility s.r.l.
  • Page 12: General Description

    GENERAL DESCRIPTION Description of the bed Bed variable height with Trendelenburg 3 joints 4 sections, electrically adjustable, top in technopolymer, shoulder side head independent. The electric bed has been designed for use in specialist or semi-intensive care units. For the purposes of the reference Directive the bed is to be understood as an active non-therapeutic device (class I).
  • Page 13 3.3.1 Title of main parts Keypad; Control Box (optional cod. 82700018); Bilateral pedals (optional cod. 82700077); Control panel Base; Telaio rete; Headboard; Footboard; Back section; 10. Basin section; 11. Upper section of the legs; 12. Lower section of the legs; 13.
  • Page 15 1. Push-button Push-button for patient and/or operator use with maintained action with possibility of deactivable controls. 2. Control Box - Control inhibitor (optional cod. 82700018); The function inhibitor panel is used when the patient’s position needs to be carefully controlled by the medical staff. The panel is fixed to the bed frame out of the patient’s reach and is used to block the functions of the free push-button.
  • Page 16: Technical Characteristics

    Technical characteristics 3.4.1 Dimensional data ELECTRIC VARIABLE HEIGHT BED 727T0045 – 727T0045C Sections Floor size 2000 x 880 Height 380 ÷ 780 Encumbrance 2260 x 1010 (960) Back section Fixed section Upper section of the legs (pelvis) Lower leg section (footboard) Backrest adjustment 0 ÷...
  • Page 17: Push Button

    PUSH BUTTON Free button panel The bed supports a free push-button with 12 buttons and 6 functions. ATTENTION! Before performing any handling, please refer to the section on the manoeuvre to be carried out. The movements that can be made by means of a push-button are as follows: Raise the back;...
  • Page 18 Inhibitor controls The electric beds can support a Control Box inhibition console, which allows operators to inhibit the functions of the free control panel. Inhibition commands disable console commands, proceed as follows: Press Tastoe A in with the padlock symbol closed and holding down the A key also press the movement key that you want to inhibitree for example: to exclude the movement of the backrest you will have to press the button A+ the button Dil green led on the edge button turns off...
  • Page 19: Control Panel( Optional Od .82700066)

    Control panel The beds can have a control system that allows the operator to control, activate and inhibit the functions of the free control panel. There are also emergency hotkeys for CPR, comfort position, Trendelenburg. and patient descent position ATTENTION! Make sure the magnetic key P is inserted. Before performing any handling, please refer to the section on the manoeuvre to be carried out.
  • Page 20 Control panels for kit 4 foldable sides ( NOT PROVIDED ) UNFORESEEN 4.5.1 Inhibition of commands from the control panel UNFORESEEN ID08435 - Rev. 1.2 - 19/04/2023...
  • Page 21: Installation

    INSTALLATION The handling activities described in this chapter shall only be carried out by qualified personnel specially trained to carry out loading operations in complete safety, unloading and handling of packages by means of lifting tools such as cranes or forklifts. Local staff should be aware of the accident prevention rules. ATTENTION! Ensure that the vehicles and logistic facilities used comply with the permitted use and in perfect condition;...
  • Page 22: Storage

    ATTENTION! During the operation, check that no part of the bed is crushed against the lifting device. Storage In case of prolonged storage, leave the bed sheltered from rain and wind and in a dry place. Especially well protect electrical parts and all parts very sensitive to humidity and low temperatures.
  • Page 23: Assembly

    Assembly ATTENTION! The assembly area must be clean and clear; must be at least 4x3 m to allow for assembly operations. The place of assembly must have the following characteristics: flat floor, not yielding; lighting 400 LUX; have an electrical power distribution socket suitable for the characteristics of the article (see identification plate) made in compliance with IEC standards.
  • Page 24: Electrical Connection

    Electrical connection WARNING! Electric beds may not be used in a potentially explosive or flammable atmosphere (hyperbaric chamber type). ATTENTION! Danger of Electrocution. Cables must be positioned so that they are not crushed, trapped, stretched, trampled, bent, wet or obstructed by moving parts. WARNING! The power cord should not interfere with the operator.
  • Page 25: Warnings

    OPERATION AND USE Warnings Electric beds can not be used in potentially explosive or flammable atmosphere (hyperbaric chamber type). Before moving the bed, make sure that the power cord is disconnected and attached to the bed. Sanitize The reader according to the modalities described in the QUICK chapter. It is the responsibility of the attending staff to authorize the patient to use the functions of the bed.
  • Page 26: Emergency Positions

    Emergency positions The bed can reach two emergency positions, depending on the type of emergency the patient is in: 1. The bed is in emergency position when the bed base is horizontally in the lowest position (total reset), with the sides lowered. 2.
  • Page 27 To bring the bed to the EMERGENCY POSITION 2 proceed as follows: press the button of the button to reset the bed sections; press the button of the push-button to bring the Treendelenburg into position; lower the sides (see reference paragraph). FROM A CONTROL PANEL press the button of the M button until the position is reached: the control clears the network plane and leads to the Trendelenburg position.
  • Page 28 To move the bed WARNING! Before moving The patient is informed. WARNING! Make sure before moving the bed that the power cord is disconnected from the mains socket and that it is properly secured so as not to hinder movement. WARNING! The handling must take place only on rigid flat surfaces and with raised sides.
  • Page 29: Alarm On Wheels For Unbraked Bed (Accessory 82700043)

    Alarm on wheels for unbraked bed (accessory 82700043) brake alarm is an alarm device for electric beds. The device alerts the operator by means of an audible alarm if the brakes are not inserted. Courtesy light device (accessory 82700086 or included in the panel kit sides 82700087) LED night courtesy light placed under the bed frame;...
  • Page 30 Raising and lowering the bed WARNING! Always notify the patient before adjusting the height of the bed. The height adjustment of the bed is achieved by two electric actuators controlled by a free control panel, a control panel, a pedal board, internal and external control panels for sides.
  • Page 31 6.9 Raising and lowering the section backrest WARNING! Always notify the patient before adjusting the back of the bed. The adjustment of the back section of the bed is obtained by means of an electric actuator controlled by free control panel, control panel and internal and external control panels for sides.
  • Page 32: Raise And Lower The

    Raise and lower the upper section of the legs WARNING! Always notify the patient before adjusting the upper leg section. The regulation of the upper section of the legs of the bed is obtained through an electric actuator controlled by free control panel, control panel and internal and external control panels for sides.
  • Page 33 7.10 Simultaneously raise and lower the backrest section and upper leg section (Autocontour) WARNING! Inform the patient before adjusting the sectionsRequire. The simultaneous adjustment of the back section and the upper section of the legs of the bed is obtained by means of two electric actuators controlled by free control panel, by control panel and by internal and external control panels for sides.
  • Page 34: Trendelenburg E Reverse Trendelenburg

    7.11 Trendelenburg and reverse trendelenburg 7.11.1 Trendelenburg WARNING! Before making any adjustments, please inform the patient. The Trendelenburg position adjustment is achieved by means of an electric actuator controlled by a free control panel, a control box, a control panel and external control panels for the sides.
  • Page 35 7.11.2 Reverse trendelenburg WARNING! Before making any adjustments, please inform the patient. The adjustment of the position of controtrendelenburg is obtained through an electric actuator controlled by free control panel, Control Box, control panel and external control panels for sides. To regulate the position of the control panel, it is necessary for the dedicated command to be enabled (see dedicated paragraphs in the chapter "CONTROL PANEL").
  • Page 36: Comfort Position (Chair)

    7.12 Comfort position (chair) WARNING! Always notify the patient before adjusting the bed sections. The comfort position (or chair) is obtained by the combined handling of the backrest section, the upper section of the legs and the inclination of countertrendelenburg. It is obtained by means of three electric actuators controlled by a control panel and internal and external control panels for sides.
  • Page 37 7.13 CPR total zero setting of positions The automatic total zeroing of the positions allows to intervene in a timely manner in emergency operations, and is obtained through electric actuators controlled by control panel and external control panels for sides and for the exclusive use of the operator.
  • Page 38: Trendelenburg Emergency

    7.14 Trendelenburg of emergency The Trendelenburg position adjustment with all sections lowered is achieved by an electric actuator controlled by a control panel and external control panels for sides and for the exclusive use of the operator. ATTENTION! The emergency switch can be activated ONLY if the override switch is in the "unlock all controls"...
  • Page 39 7.15 Emergency release device back CPR The CPR backrest release device allows the backrest to be lowered quickly in case of emergency and can be activated by means of a bilateral lever positioned under the bed base. To lower the backrest proceed as follows: ▪...
  • Page 40 7.16 Security position at night The bed is equipped with a "safety night" position, which is a function that automatically brings the bed to the minimum height of 380 mm from the floor. This position is obtained by means of an electric actuator controlled by a magnetic key control panel, a control panel and external control panels for the exclusive use of the operator.
  • Page 41 7.17 Raise and lower the lower leg section WARNING! Before adjusting the lower part of the patient’s blood pressure, check the patient. The lower section of the legs can be adjusted by two "Rastomat" 6-position trigger mechanisms or by an electric actuator (optional). WARNING! Do not insert hands or objects between the lower section of the legs and the bed base or the sides.
  • Page 42: Integrated Bed Extension ( Optional Code 82700038 )

    7.19 Integrated bed extension (optional cod. 82700038) WARNING! The patient should not have to wait for the duration of this regulation. Built-in bed extension to the bed frame. It allows the extension of the bed top of 18cm . Made with a steel profiled frame and protected by a cover in technopolymer. To stretch the bed proceed as follows: make sure the bed is locked (see par."Lock and unlock the bed");...
  • Page 43 7.20 Handling 4 bed areas 727T0045 WARNING! Always notify the patient before making any adjustments. To lower the sides proceed as follows: grab the bank with one hand and with the other lower the lever in the direction indicated by the arrow (15); lower the bank by turning it downwards.
  • Page 44 7.21 Handling of bed hinges cod.727T0045C WARNING! Always notify the patient before making any adjustments AVVERTENZA! Prima di effettuare la manovra avvisare sempre il paziente. ATTENZIONE! Prima di effettuare qualsiasi manovra assicurarsi che il freno sia inserito. Il letto è dotato di sponde utili a contenere il paziente riducendo il rischio di cadute accidentali. Le sponde sono facilmente abbattibili in modo da consentire un agevole ingresso/uscita dal letto.
  • Page 45 7.22 Disassemble the sides (cod. 7300007A) To disassemble the sides with compass proceed as follows: press the release lever both front and back, downwards; Pull the tailgate pivot arms outwards until the clamp pins have left the seat; repeat for the other bank; place the sides in a safe place.
  • Page 46 7.23 Equipotential connection The electric beds are equipped with equipotential connecting clamp placed on the side of the bed head; the clamp is necessary for the equalization of the electrical potentials of all the metal parts without protection. WARNING! DANGER OF ELECTRICAL DEVIATIONS. It is necessaryto usefully equipotential connection cauldron if the patient is connected to intravascular or intracardiac equipment.
  • Page 47 7. OPTIONAL Cod.82700002 – ROD LIFTS PATIENT WITH TRAPEZE Patient lifting rod made of chromed steel tube. Adjustable strap and molded trapeze. Cod. 82700025 - HEIGHT ADJUSTABLE DRIP ROD 4 HOOKS IV rod adjustable in height to 4 hooks. Cod. 82700016 - HEIGHT ADJUSTABLE DRIP ROD 4 STAINLESS STEEL HOOKS Height adjustable IV rod with 4 stainless steel hooks.
  • Page 48: Sanitation

    SANITATION Sanitizing products ATTENTION! Sanitizing agents are corrosive. The best sanitizing agents are the most commonly used in industry. Follow the instructions provided by the manufacturer for the specific application during use. If possible, ask the manufacturer for guarantees on the degree of corrosivity of the solutions used. Any change to these features may damage the item.
  • Page 49: Maintenance

    Pam Mobility s.r.l. Via Verdi 39 - 42043 Gattatico (RE) - Italy Tel 0522 473859 - Fax 0522 1548244 E-mail: info@pammobility.comhttp: www.pammobility.com Specifying: ▪ Product code, serial number, production code, year of installation; ▪ defects found; ▪ exact address of the place where the bed is installed.
  • Page 50: Battery

    Demolition and disposal The materials of the bed consist essentially of: ▪ painted or galvanized ferritic steel; ▪ plastic material in abs; ▪ elastomers; Disassemble the bed separating the individual pieces according to the material with which they are made, it is mandatory to dispose of the different materials in accordance with the regulations of the country in which the bed must be removed.
  • Page 51: 10. Guarantee

    Troubleshooting DEFECT POSSIBLE CAUSE INTERVENTION Not connected to the power line. Connect to the power line. If interchangeable cable replace with new. If The power supply device does not turn on Power cord is defective. fixed cable ship to service center. It’s a faulty control unit.
  • Page 52 11. WIRING DIAGRAM PUSH BUTTON INPUT 1 ENGINE BACK (f) INPUT 2 ENGINE DOCK (g) INPUT 3 ENGINE RISE (c) INPUT 4 ENGINE RISE (c) INPUT 5 FOOT MOTOR (h) (where applicable) POS. DESCRIPTION Control unit TC21 Control unit MCL care 4 outputs 230V IP 66 Buffer battery TBB2-4398-001-0 ACO TNP6-4398-001 Push-button panel TH12...
  • Page 53 Via Verdi, 39 - 42043 Gattatico (RE) Tel. +39 0522 473859 - Fax +39 0522 1548244 info@pammobility.com- www.pammobility.com...

This manual is also suitable for:

727t0045727t0045c727t0045s