Manufacturer's Recommendations
15 Manufacturer's Recommendations
UNIT:
MANUFACTURER:
Instrument has to undergo a safety regulation control every 24 months.
E X T E N T:
1.
Check housing, power cable, electrode connection sockets, electrode cables, plate electrodes,
diplodes, diodes and rubber capacitor electrodes for deformations and damages.
2.
Correct function of indicators
3.
Correct display of operating modes
4.
Accuracy check of timer (+/- 5s), time setting
5.
Correct function of power switch
6.
Verification of power output using a neon indicator
7.
Electrical test according to IEC 601-1
7.1 Earth leakage current
7.2 Housing leakage current
7.3 Protective conductor resistance
52
M A N U F A C T U R E R ' S R E C O M M E N D A T I O N S
S A F E T Y R E G U L A T I O N S C O N T R O L
according to Medical Devices Directive
PHYSIOTHERM-S
PHYSIOMED ELEKTROMEDIZIN AG
normal condition
≤ 0.5 mA
≤ 0.1 mA
100 mOhm
single fault condition
≤ 1.0 mA
≤ 0.5 mA
PHYSIOTHERM-S