Bed Location - PROMA REHA VERTICO Instructions For Use Manual

Hospital bed with verticalization for standard care
Table of Contents

Advertisement

Parameters of electrical parts used on the bed, Onimed
Operating voltage
Supply frequency
Maximum input
Protection class
Applied part
Protection
Motor load
Nominal value of fuse
Weighted noise power level
B
L
ED
OCATION
The bed must be placed in such a way to maintain at least 10 cm free space around it on each side and at least 20 cm from the wall
or surrounding objects on the side of the headboard. Furthermore, make sure that no items, such as shelves, televisions, etc., are
placed above the bed, which could be released or pulled down by the bed when adjusting its position or lifting it and it could therefore
result in the injury to the operator or the patient! If the bed is equipped with accessories (e.g. trapeze bar, infusion stand, extension
kit, etc.), the operator controlling the bed must ensure that none of the parts of the bed or accessories touches, leans against or
otherwise contacts with the ceiling, niches or other fixed or removable objects placed on the walls or dropped ceilings, such as
shelves, lamps, etc. If there is not enough space for such handling, remove the accessories or place the bed on another place, so that
it does not come into contact with other objects.
If the bed is equipped with tilting of the bearing area to the vertical position, it must always be located to avoid the collision with the
surrounding objects in any of these positions. No accessories may be installed when using these positions. The trapeze bar with a
small vertical bar or the infusion stand of the above-mentioned extreme positions exceed the contour of the bed. They could cause
damage to the surrounding objects or injury to a third party!
Always place the bed in such a way that there is at least 5 cm free space on each side of it. When adjusting the bearing
area height, it may be deflected. If the bed is placed close to the surrounding objects (walls), it could be damaged, or
it could damage the surrounding things or cause bodily harm of the patient, the operator or a third party!
If the bed is equipped with the verticalization function, place it always in such a way that it does not collide with the
surrounding objects in both extreme positions!
If the bed is fitted with the accessories (trapeze bar with a small vertical bar, an infusion stand, etc.), observe the
accessories trajectory each time the bed is adjusted!
Before the bearing area verticalization, always remove the accessories, otherwise they may fall out!
Instructions for Use of VERTICO Bed
230 V
50/60 Hz
350 W
I
B type
minimum IPS4
10%, 2 min. max. load/18min. At rest
5 A
48 dB measured according to ISO 3746, not loaded
Page 8/32