10.2.1 FUNCTION LOCK OF THE PATIENT CONTROLLER

Individual functions of the patient controller can be locked. Locked function is indicated by
changing the illumination from white to red for the given function.
Lock the individual functions using the magnetic wrench as follows:
Activate the controller using the GO button (1)
Attach the magnetic key to the lock (see figure); the buttons on the
controller start flashing (lock mode indication)
Press the button of the function you want to lock.
For the given function, the illumination changes from white to red
Terminate the lock mode by pressing the GO button (1); the
buttons on the controller will stop flashing
UNLOCK the function by repeating the above procedure. After pushing the locked function,
the function lock indication will go out (10).
After the universal plastic key has been inserted into the controller, the buttons on
the patient controller will start flashing!
The function locks to improve patient safety. Professional health staff decide on
the patient's ability to perform partial positioning of the bed separately. Always
carefully consider which functions can be safe and available for self-control by the
patient, taking into account their health and psychological condition!
One universal key can be used to lock all beds of the same type!
10.3 S
B
TANDBY
ATTERIES
As a standard, the beds are equipped with batteries, which are used for emergency operation of the bed in case of power failure. If
the bed is connected to the electricity network, this connection is indicated by the Power connection icon (11) in the top right corner
of the nurse panel or at the bottom of the patient hand controller. The battery is charged automatically when connected to the
electricity network. The battery charge status is indicated by a Battery status display (12). The following statuses can be detected
by the activation of the LEDs:
...
...
...
...
...
...
...
...
The bed is not primarily intended for operation from the backup battery. Operate the bed from backup batteries in
exceptional cases (power outage, transport) only. This will prolong the life of backup accumulators!
Instructions for Use of VERTICO Bed
Bed connected to the network
Battery charging (Full battery charge min. 10
hours)
Battery charging finished.
The bed is connected to the electricity network.
Critical battery condition – battery not functional or disconnected. Contact the service centre.
The battery is in good condition charged
Low battery level. The battery is charged to
less than 25% of its capacity. Connect the
bed to the electricity network! (min. 10
The bed is disconnected from the electricity
Page 14/32
Bed disconnected from the
network
more than 25% of its capacity.
hours)
network.