Before moving the bed itself, always adjust it to the basic position (i.e. all the adjustable parts are in the lowest possible position).
Set the height of the bearing area to the optimum position, but always not more than 20 cm below the maximum reachable
height. When relocating and handling, the guards must always be secured in the upper position to reduce the risk of a patient
fall. Disconnect the power supply cable, hang it on the headboard and secure the fork to prevent the cable or fork moving or pinching
during the transport of the bed. Similarly, secure the nurse controller to prevent it or its power supply cable from damaging. Make
sure that the power supply cable or controller is not pulled over the ground. This could damage them, break them or damage the
insulation. When reconnecting the bed, follow the Instructions for Use, especially observe the points listed in the chapter Installation
and Commissioning of the Bed.
The bed can be moved in the basic position only (i.e. the bearing area and all the adjustable parts are in the lowest
possible position)!
Set the height of the bearing area to the optimum position, but always not more than 20 cm below the maximum
reachable height.
When relocating and handling, the guards must always be secured in the upper position to reduce the risk of a patient
fall.
The bed may only be moved with a properly secured power supply cable and a controller against damage!
9.1
S
O
WITCHING
FF
If necessary to switch off the bed during transfer or storage, it is possible to switch it to the transport/storage mode (off):
Disconnect the power supply cable of the bed from the electricity network (1)
Activate the controller using the GO button (2)
Press the GO and STOP buttons simultaneously and hold them for at least 5 seconds (3) - the entire electronics
will switch off and the controller will go out
1
The electronics can be switched on again by connecting the bed to the electricity network or by short pressing (<1 second) the reset
button on the control box.
9.2
S
O
WITCHING
N THE
You will switch the bed on by connecting the power supply cable to the network or by short pressing the RESET button on the control
box.
Instructions for Use of VERTICO Bed
H
ANDLING
-
THE BED WITH THE PATIENT CONTROLLER
2
B
ED
4 (5 seconds)