Clamp For Electrical Potential Balancing; Power Supply Cable Holder; Mechanical Control Elements; Central Brake - PROMA REHA VERTICO Instructions For Use Manual

Hospital bed with verticalization for standard care
Table of Contents

Advertisement

10.4
C
E
LAMP FOR
The bed may be equipped with a terminal for balancing the electrical potential (1) which is located in the head section of the bed
frame.
Use the clamp when connecting the bed to the medical isolated system, where it serves as a grounding system, or when scanning
electrical potentials, it is used to balance the potentials between the patient and the device. The connected clamp allows balancing
the potential of all conductive parts of the bed and other connected electrical devices.
Interconnect the clamp with devices by connecting to a standardized medical terminal located on the
bearing surface frame in the bed head section. Place the electric potential balancing conductor (2)
on this terminal with the corresponding standardized connector .
The clamp for balancing the electrical potential is marked on the bed with a pictogram:
Only the conductor of the potential connection may be connected to the clamp for
balancing electrical potential. NO VOLTAGE MUST be connected to the clamp!
Before transporting the bed, disconnect the conductor of electrical potential balancing
from the bed, otherwise it might result in conductor breaking or damage to the bed!
10.5
P
S
OWER
UPPLY
The bed is equipped with a power supply cable holder (1)attached to the power supply cable. Read the instructions in chapter 9
before starting any transport. RELOCATION AND HANDLING
Disconnect the power supply cable from the electricity network
Hang the plastic holder (1) on the headboard (or frame) (2)
Wind the power supply cable on the power supply cable holder
Secure the fork and cable so that the cable or fork cannot be
moved or pinched during the bed transport
The bed may only be moved with properly secured power supply cable against damage! The transport of the bed
with the unsecured power supply cable may result in its serious damage and subsequent increased risk of electrical
injury!
M
ECHANICAL
11.1 C
B
ENTRAL
RAKE
The bed chassis is equipped with four casters with a central brake. From the patient's point of view under the head to the right, the
arresting wheel is always placed, which may be replaced by the fifth arresting wheel in the bed centre upon the customer's request.
The arresting wheel is used to facilitate handling of the bed during transport.
The central brake lever can be set to three positions:
POSITION 1 – all casters are braked
POSITION 2 (horizontal position) – all casters are released
POSITION 3 – the casters are released and the arresting or fifth wheel is activated
for easier handling during transport
Always operate the central brake lever using the lower limb tip. When using
a different mode of control, there may occur a risk of bodily harm by
squeezing!
The bed must always be properly braked apart from its transport. Medical
staff must always make sure that the bed is braked before use. If the bed is
not braked, there may exist an increased risk of injury, especially when
standing from a sitting position, seating on the bed or leaning against it!
Instructions for Use of VERTICO Bed
P
LECTRICAL
OTENTIAL
C
H
ABLE
OLDER
C
E
ONTROL
LEMENTS
B
ALANCING
Page 15/32