14.4 E
M
RROR
ESSAGE
The bed is provided with electronics of own diagnostics. Use a light code to determine an error and use the Instructions for Use to
try to remove it. The light code is displayed on the manual patient controller in the form of flashing red LED during positioning. The
displayed code always appears when positioning the relevant problem part and lasts until a cause of the error message is removed.
Light code
flashes
briefly 2x
Motor/s overloaded
● ●
● ●
● ●
flashes
briefly 3x
● ● ●
● ● ●
Briefly 1x and
long flashing 1x
flashing briefly
IRC Sensor Error
● ● ● ● ● ● ● ● ●
14.4.1

MOTOR OVERLOADED

If the bed diagnostics signals this error message, continue with the following steps:
1)
Stop positioning
2)
Press the GO button to deactivate signalling
3)
Remove obstacles that may prevent the relevant motor from moving (decrease engine load)
4)
Try again to adjust shortly the position of the relevant part. If the error message does not appear, you can continue
in positioning. If the error message is displayed again, contact the authorised service department, the motor is damaged.
14.4.2

MOTOR NOT CONNECTED

If the bed diagnostics signals this error message, continue with the following steps:
1)
Stop positioning
2)
Press the GO button to deactivate signalling
3)
Check the connection of the connectors to the control box (the cables pulled out a bit) and the
relevant cable for mechanical damage. If you do not find any mechanical damage to the cabling,
continue. If the cable is damaged mechanically, disconnect the bed from the electricity network,
switch off the bed and contact the authorized service centre. Do not change the bed position any
more. There is a risk of electric shock!
4)
Try again to adjust shortly the position of the relevant part. If the error message does not appear, you can continue
in positioning. If the error message is displayed again, contact the authorised service department, the motor is damaged.
Instructions for Use of VERTICO Bed
THE DIODES FLASH ON THE CONTROLLER
Fault
* Bed maximum capacity was exceeded for a short term
* The motor is damaged and is short
* Motor or gear mechanics is blocked
* Motor connector is not sufficiently plugged in the control box
Motor/s not
* Motor cable damaged
connected
* Motor damaged
* The tolerance of the difference in a position of the individual column motors
Position errors
was exceeded
* Motor position sensor damaged
Possible cause
Page 24/32