P
ROBLEMS AND
14.1 T
HE BUTTONS ON THE
If the buttons on the patient hand controller are flashing, the lock mode is activated. To use the standard control, remove the universal
plastic key from the controller, see chapter 10.3.1. FUNCTION LOCK OF THE PATIENT CONTROLLER
14.2 N
-F
ON
UNCTIONAL
If the bed cannot be controlled via the controllers and the electric power supply indication diode on the nurse panel is not illuminated,
try to repair the bed using the following procedure.
Check the bed for connecting to the electricity network. If the bed is not connected, insert the power supply cable into
the electricity network. Then the indication diode on the nurse panel goes on and the bed can be fully controlled. If not, turn
on the bed using the RESET button on the control box. If the bed still does not respond, continue.
Disconnect the power supply cable from the electricity network and check it for damage. If it is not damaged, insert
the power cable into the electricity network and continue. If you find damage, you must not use the bed any longer and
you must contact the service centre!
Check, whether the electrical network for connecting the bed is working. If not, ensure that the network is put in
operation. Then the indication diode on the nurse panel goes on and the bed can be fully controlled. If not, continue.
Check whether the green diode on the control unit indicating supply from the electricity network is on. If this diode
does not emit light, disconnect the bed from the electricity network and contact the service centre. If the diode is on, continue.
Disconnect the bed from the electricity network and check the cables of the control elements (patient controller) for
damage. If you find any damage, you must not use the bed any longer and you must contact the service centre. If no damage
is found, contact the service centre.
14.3 N
-F
ON
UNCTIONAL
If an unexpected error occurs, the bed happens to emit light but does not respond to any control. In this case, restart the bed. After
resetting, always check the bed function. You can reset software and switch off the bed using the button on the control box.
Functions
Bed reset
Instructions for Use of VERTICO Bed
T
ROUBLESHOOTING
P
H
ATIENT
AND
B
ED
DOES NOT EMIT LIGHT
B
ED
EMITS LIGHT
Button on the control box
Shortly press and release the reset button located on the control box.
Hold the button for less than 1 second
.
C
F
ONTROLLER ARE
LASHING
,
DOES NOT RESPOND TO THE CONTROL
,
DOES NOT RESPOND TO THE CONTROL
Task
Page 23/32
Description
Reset the internal bed
programme