► In case of changes in or loss of functionality, discontinue use of the
product and have it checked by authorised, qualified personnel.
CAUTION
Excessive strain due to use on more than one patient
Risk of injury and loss of functionality as well as damage to the product
► Use the product on only one patient.
► Observe the maintenance recommendations.
NOTICE
Exposure of the product to unsuitable environmental conditions
Damage, brittleness or destruction due to improper handling
► Avoid storage in condensing ambient humidity.
► Avoid contact with abrasive substances (e.g. sand, dust).
► Do not expose the product to temperatures below -10 °C (14 °F) or
above +60 °C (140 °F) (e.g. sauna, excessive sunlight, drying on a
radiator).
5 Preparation for use
5.1 Processing
Required materials: Alignment adapter no. 4 (743Y56=4), lamination
dummy with screw (17B206=*), 2 lamination bars 16 mm (e.g.  17LS3=16-T)
>
The plaster model was produced specifically for the patient.
1) Position the lamination dummy on the plaster positive, on the medial side
at the compromise pivot point according to Nietert.
2) Secure the lamination bars on the lamination dummy using the screws
and check the alignment on the model.
3) Mark the insertion range on the lamination bar.
4) Loosen the screws of the lamination bars and adapt them with bending
irons.
5) Secure the lamination bars on the lamination dummy using the screws.
6) Repeat the steps until the desired result is obtained.
7) Insert the shoulder screw through the lamination dummy and affix it to
the alignment adapter.
8) Laminate the orthosis to fit the patient.
9) Cut free and remove the lamination dummy.
5.2 Assembly
The medial support is pre-assembled on delivery.