Hantek Tablet1000 Series User Manual

Hantek Tablet1000 Series User Manual

Tablet oscilloscope
Hide thumbs Also See for Tablet1000 Series:
Table of Contents

Advertisement

Tablet1000
series
Tablet oscilloscope
User Manual
2022.05

Advertisement

Table of Contents
loading

Summary of Contents for Hantek Tablet1000 Series

  • Page 1 Tablet1000 series Tablet oscilloscope User Manual 2022.05...
  • Page 2 Hantek certified Tablet1000 series oscilloscope to meet China's national industry standards and has passed the CE certification. Contact us If you have any questions when using the products of Qingdao Hantek Electronics Co., LTD., you can obtain service and support through the following ways: Email: service@hantek.com, support@hantek.com...
  • Page 3: Table Of Contents

    Document overview ....................8 Quick start ......................11 General examination ..................11 Appearance and dimension ................11 Preparation before use ..................12 4.3.1 Connect the power supply ....................12 4.3.2 Boot check ........................13 Copyright © Qingdao Hantek Electronics Co., LTD Tablet1000 User manual...
  • Page 4 Adjust the horizontal time base (time/grid) ............24 Pan or zoom a single collection or stopped collection ........25 Set the vertical system ..................26 Open or close the analog channel ..............26 Tablet1000 User manual Copyright © Qingdao Hantek Electronics Co., LTD...
  • Page 5 Trigger oscilloscope .................... 37 Trigger level .....................37 Trigger data source ..................37 Trigger mode ....................38 Trigger hold-off ....................38 Trigger type ......................39 9.5.1 Edge-triggered ........................ 39 9.5.2 Pulse width trigger ......................40 Copyright © Qingdao Hantek Electronics Co., LTD Tablet1000 User manual...
  • Page 6 Vertical Cursor ........................ 50 10.3.2 Horizontal Cursor ......................51 11 Acquisition ......................53 11.1 Acquisition mode ....................53 11.2 Average times ....................53 11.3 Display mode ....................54 11.4 Storage depth ....................55 11.5 Interpolation methods ..................56 Tablet1000 User manual Copyright © Qingdao Hantek Electronics Co., LTD...
  • Page 7 17 Signal source ...................... 68 17.1 Set the output waveform type and parameters ..........68 17.2 Set waveform modulation .................69 17.3 Set burst ......................69 17.4 Edit arbitrary waveform ..................70 17.5 Output arbitrary waveform ................71 Copyright © Qingdao Hantek Electronics Co., LTD Tablet1000 User manual...
  • Page 8 19 Settings ....................... 75 20 Gallery......................... 78 21 Remote control ....................79 22 Fault handling ..................... 81 23 Appendix ......................82 23.1 Appendix A: Accessories ..................82 23.2 Appendix B: Warranty summary ...............84 Tablet1000 User manual Copyright © Qingdao Hantek Electronics Co., LTD...
  • Page 9: Figures

    Figure 6.1 Channel vertical menu ............... 26 Figure 6.2 Waveform inversion ................29 Figure 7.1 Math operation menu ................. 31 Figure 7.2 Addition operation ................32 Figure 7.3 FFT operation ..................33 Copyright © Qingdao Hantek Electronics Co., LTD Tablet1000 User manual...
  • Page 10 Figure 16.1 Multimeter ..................65 Figure 17.1 Signal source ................... 68 Figure 17.2 Arbitrary wave editor ................ 70 Figure 18.1 File browse ..................73 Figure 20.1 Gallery ..................... 78 Tablet1000 User manual Copyright © Qingdao Hantek Electronics Co., LTD VIII...
  • Page 11 Table 3.1 Model ....................10 Table 15.1 Automatic setting ................63 Table 15.2 Automatic setting of sine waves ............63 Table 15.3 Default setting ................... 64 Table 19.1 Settings ..................... 75 Copyright © Qingdao Hantek Electronics Co., LTD Tablet1000 User manual...
  • Page 13: Safety Requirem Ents

    Do not touch exposed connectors and components after power is switched on. ⚫ Do not operate if the product is suspected to be faulty. If you suspect that the product has been damaged, please ask qualified maintenance personnel to check it. Copyright © Qingdao Hantek Electronics Co., LTD Tablet1000 User manual...
  • Page 14: Security Terms And Signs

    Safety terms on products: Warning: Indicates a potential hazard may be caused to you if you do not perform this operation. Safety signs on the product: Measurement category Measurement category Tablet1000 User manual Copyright © Qingdao Hantek Electronics Co., LTD...
  • Page 15: Working Conditions

    In and out of operation: 3,000m (10,000 ft). Random vibration: 0.31g RMS at 50Hz to 500Hz, 10 minutes in each axial direction. No operation: 2.46g RMS from 5Hz to 500Hz, 10 minutes in each axial direction. Copyright © Qingdao Hantek Electronics Co., LTD Tablet1000 User manual...
  • Page 16: Maintenance And Cleaning

    Do not expose the LCD to direct sunlight for a long time when storing or placing the oscilloscope. Cleaning: If the oscilloscope and probe are inspected frequently as required by operating Tablet1000 User manual Copyright © Qingdao Hantek Electronics Co., LTD...
  • Page 17: Environmental Considerations

    Copyright © Qingdao Hantek Electronics Co., LTD Tablet1000 User manual...
  • Page 18: Product Features

    4 analog channels or 2 analog channels;Its own signal source support 25MHz standard waveform output, built-in a variety of arbitrary wave;The Tablet1000 User manual Copyright © Qingdao Hantek Electronics Co., LTD...
  • Page 19 Standard 9V2A/5V3A/ 12v1.5A charger, USB interface input, can share a set of charging equipment with mobile phone; Powered by 4 or 2 18650 lithium batteries; Adopt the design of supporting foot, as you use as you put. Copyright © Qingdao Hantek Electronics Co., LTD Tablet1000 User manual...
  • Page 20: Document Overview

    Acquire menu. Document content conventions: Tablet1000 series tablet oscilloscope consists of the following models. Unless otherwise specified, this manual uses TO1254D as an example to describe the Tablet1000 series and basic operations. Model Channel Bandwidth Sampling rate...
  • Page 21 1GSa/S TO1154C 150MHz 1GSa/S TO1204C 200MHz 1GSa/S TO1254C 250MHz 1GSa/S TO1154D 150MHz 1GSa/S 25MHz TO1204D 200MHz 1GSa/S 25MHz TO1254D 250MHz 1GSa/S 25MHz TO1154AUTO 150MHz 1GSa/S 25MHz TO1204AUTO 200MHz 1GSa/S 25MHz Copyright © Qingdao Hantek Electronics Co., LTD Tablet1000 User manual...
  • Page 22: Table 3.1 Model

    Document overview Model Channel Bandwidth Sampling rate Signal source Multimeter TO1254AUTO 250MHz 1GSa/S 25MHz Table 3.1 Model Tablet1000 User manual Copyright © Qingdao Hantek Electronics Co., LTD...
  • Page 23: Quick Start

    If you find the instrument is damaged, not working properly, or unable to pass the performance test, please contact the dealer responsible for this business. Appearance and dimension Figure 4.1 Front view4 Copyright © Qingdao Hantek Electronics Co., LTD Tablet1000 User manual...
  • Page 24: Preparation Before Use

    When charging, if the machine is not started, the indicator light of the power button will be on. If the machine is turned on, the battery box on the screen will change (slow charge , quick charge Tablet1000 User manual Copyright © Qingdao Hantek Electronics Co., LTD...
  • Page 25: Boot Check

    Do not touch the metal part on the top of the probe when the probe is connected to the high voltage power supply. Before any measurement, connect the probe to the oscilloscope and ground the ground end. Copyright © Qingdao Hantek Electronics Co., LTD Tablet1000 User manual...
  • Page 26: Figure 4.3 Probe

    Figure 4.4 Probe compensation4 3) If necessary, adjust the variable capacitance on the probe with a non-metallic screwdriver until the waveform displayed on the screen is "compensated correctly" as shown above. Tablet1000 User manual Copyright © Qingdao Hantek Electronics Co., LTD...
  • Page 27: Product Introduction

    This section introduces the front, rear, side panel and the main interface of the oscilloscope. 4.4.1 Front Panel Figure 4.6 Front panel Label Hantek trademark + product model. Virtual keys Copyright © Qingdao Hantek Electronics Co., LTD Tablet1000 User manual...
  • Page 28: Rear Panel

    [PRINT] : Screenshot key. Save the current screen to internal memory or external memory. 800*480 resolution LCD. 4.4.2 Rear Panel Figure 4.7 Rear view Stand With the hanging supporting stand, it can be used anywhere. Tablet1000 User manual Copyright © Qingdao Hantek Electronics Co., LTD...
  • Page 29: Side Panel

    Power switch When the machine is off, press the power switch to start the machine. Press the power switch again to shut down the machine. Figure 4.9 Left panel USB interface Copyright © Qingdao Hantek Electronics Co., LTD Tablet1000 User manual...
  • Page 30: Figure 4.10 Upper Panel

    1KHz square wave signals. CH1 ~ CH4 channels 4 analog channels that connects the probe and analogs signal input. Tablet1000 User manual Copyright © Qingdao Hantek Electronics Co., LTD...
  • Page 31: Main Interface

    Tap to enter the gallery page. Pictures saved by PRINT and saved in the oscilloscope interface will be displayed in the file browsing. Battery Display the battery level in real time. Copyright © Qingdao Hantek Electronics Co., LTD Tablet1000 User manual...
  • Page 32: Touch Screen Gesture

    Move two fingers together or apart. Kneading gestures allow you to zoom in and out of the waveform. Slide the close fingers apart to enlarge a waveform and slide the separated fingers closer together to shrink a waveform as the following figure shows. Tablet1000 User manual Copyright © Qingdao Hantek Electronics Co., LTD...
  • Page 33: Dragging

    Click the parameter to be set, and the virtual keyboard will pop up. Set parameters through the keyboard. Input a string When naming a file or folder, you need to enter a string as shown in the following figure. Copyright © Qingdao Hantek Electronics Co., LTD Tablet1000 User manual...
  • Page 34: Figure 4.15 Character Keyboard

    After input, click on the "Enter" button. Input numerical values When modifying the parameters, the numeric keypad is used to input the numerical value as shown in the following figure. Tablet1000 User manual Copyright © Qingdao Hantek Electronics Co., LTD...
  • Page 35: Figure 4.16 Numeric Keypad

    Increase or decrease the number. Left and right arrows to determine the number, up and down arrows to increase or decrease the number. Long press the up and down arrows to change the number continuously. Copyright © Qingdao Hantek Electronics Co., LTD Tablet1000 User manual...
  • Page 36: Set The Horizontal System

    The horizontal time base is the time value represented by each grid in the horizontal direction of the display screen, generally expressed as s/div. If the horizontal time base is changed, the horizontal time base will be displayed in real time as shown below. Tablet1000 User manual Copyright © Qingdao Hantek Electronics Co., LTD...
  • Page 37: Pan Or Zoom A Single Collection Or Stopped Collection

    After the oscilloscope stops, the display screen may contain several useful collection data, but you can only pan and zoom the last collected data. Pan and zoom data from a single collection or stopped collection. Copyright © Qingdao Hantek Electronics Co., LTD Tablet1000 User manual...
  • Page 38: Set The Vertical System

    If the channel label is in yellow as shown below, it indicates that the channel is open. If the channel label is gray, it indicates that the channel is closed. Tablet1000 User manual Copyright © Qingdao Hantek Electronics Co., LTD...
  • Page 39: Adjust The Vertical Sensitivity

    Use the touch screen function to adjust the vertical offset by dragging gestures. For details, see the Dragging section. In the Vertical channel menu, increase or decrease the offset by tapping the arrow icon Copyright © Qingdao Hantek Electronics Co., LTD Tablet1000 User manual...
  • Page 40: Channel Coupling

    After it is enabled, the color of B on the channel label is the same as that of the channel as shown in the figure below. Tablet1000 User manual Copyright © Qingdao Hantek Electronics Co., LTD...
  • Page 41: Waveform Inversion

    Amplitude unit Open Vertical channel > Unit. The optional units are voltage and current. The default unit is voltage. Copyright © Qingdao Hantek Electronics Co., LTD Tablet1000 User manual...
  • Page 42 Set the vertical system When the amplitude unit is changed, the unit of related functions are also changed. Tablet1000 User manual Copyright © Qingdao Hantek Electronics Co., LTD...
  • Page 43: Mathematical Operations

    Data source 1: Select data source CH1 to CH4 to perform the operation. Data source 2: Select data source CH1 to CH4 to perform the operation. Algebraic operations, taking addition as an example: Copyright © Qingdao Hantek Electronics Co., LTD Tablet1000 User manual...
  • Page 44: Fft Operation

    Bartlett: Good frequency resolution. Used to measure strong signals with narrow bandwidth. Flat top: Good amplitude resolution, poor frequency resolution. Used to measure Tablet1000 User manual Copyright © Qingdao Hantek Electronics Co., LTD...
  • Page 45: Figure 7.3 Fft Operation

    To reduce the DC component, set the "Channel coupling" of the data source to "AC" mode. In order to reduce the random noise and aliasing frequency component of repeated or Copyright © Qingdao Hantek Electronics Co., LTD Tablet1000 User manual...
  • Page 46 Use the AX and BX cursors to measure the frequency and the difference between the two frequencies (BX-AX). Use the AY and BY cursors to measure the amplitude and amplitude difference (BY-AY). For details, see the Cursor measurement section. Tablet1000 User manual Copyright © Qingdao Hantek Electronics Co., LTD...
  • Page 47: Reference Channel

    Click to select a reference waveform file, and a pop-up box will prompt: Call out successfully. A gray reference waveform will appear on the screen. Click the yellow bar , it will turn gray, indicating the reference waveform is closed. Copyright © Qingdao Hantek Electronics Co., LTD Tablet1000 User manual...
  • Page 48: Figure 8.1 Reference Waveform

    The method of adjusting the horizontal or vertical position of the reference waveform is the same as adjusting the channel. For details, refer to Set the horizontal system the vertical system. Tablet1000 User manual Copyright © Qingdao Hantek Electronics Co., LTD...
  • Page 49: Trigger Oscilloscope

    Trigger data source Open Trigger > Source and select the data source. Available data sources are CH1 ~ CH4. The selected channel works whether it is opened or not. Copyright © Qingdao Hantek Electronics Co., LTD Tablet1000 User manual...
  • Page 50: Trigger Mode

    The hold-off can be triggered on repeated waveforms that have multiple edges (or other events) between themselves. If the minimum time between triggers is known, you can also use hold-off to trigger on the first edge of the trigger. Tablet1000 User manual Copyright © Qingdao Hantek Electronics Co., LTD...
  • Page 51: Trigger Type

    Click Source and select CH1 ~ CH4 as the trigger source. CH1 to CH4: analog channel. Click Mode and select the trigger mode (automatic, normal, single). Copyright © Qingdao Hantek Electronics Co., LTD Tablet1000 User manual...
  • Page 52: Pulse Width Trigger

    (positive and negative). Positive: Triggered when a signal with positive polarity pulse width occurs. Negative: Triggered when a signal with negative polarity pulse width occurs. Click When and select the trigger condition. Tablet1000 User manual Copyright © Qingdao Hantek Electronics Co., LTD...
  • Page 53: Figure 9.3 Pulse Width Trigger Menu

    <, > : The oscilloscope will be triggered when the pulse width of the data source signal is less than or greater than the specified pulse width. Figure 9.3 Pulse width trigger menu Copyright © Qingdao Hantek Electronics Co., LTD Tablet1000 User manual...
  • Page 54: Video Trigger

    As shown in the figure below, the time difference between the two points (A and B) where the high and low trigger levels intersect the rising (falling) edge of the waveform is defined as the positive (negative) slope time. Tablet1000 User manual Copyright © Qingdao Hantek Electronics Co., LTD...
  • Page 55: Timeout Trigger

    Trigger when the time interval (△T) from the rising edge (or falling edge) of the input signal passing through the trigger level to the adjacent falling edge (or rising edge) passing through the trigger level is greater than the set timeout time. As shown below: Copyright © Qingdao Hantek Electronics Co., LTD Tablet1000 User manual...
  • Page 56: Figure 9.6 Timeout Trigger Menu

    Negative polarity: Start timing when the falling edge of the input signal passes through the trigger level. Click Time to set a reference value for the timeout period. Figure 9.6 Timeout trigger menu Tablet1000 User manual Copyright © Qingdao Hantek Electronics Co., LTD...
  • Page 57: Measure

    ⚫ To cancel the measurement parameter, click the measurement item again, or click “Clear all”. Figure 10.1 Measurement parameters Copyright © Qingdao Hantek Electronics Co., LTD Tablet1000 User manual...
  • Page 58: Time Parameters

    10. Maximum time: The time value corresponding to the maximum waveform (Vmax). 11. Minimum time: The time value corresponding to the minimum waveform (Vmin). Note: The upper threshold, middle threshold and lower threshold are 90%, 50% and 10% respectively. Tablet1000 User manual Copyright © Qingdao Hantek Electronics Co., LTD...
  • Page 59: Count Parameters

    Trigger count: The number of rising edges (or falling edges). Note: The above measurements only apply to analog channels. The upper and lower thresholds are 90% and 10% respectively. Copyright © Qingdao Hantek Electronics Co., LTD Tablet1000 User manual...
  • Page 60: Delay And Phase Parameters

    "delay 1->2" and data source 1, expressed in degrees. The phase calculation formula is: Phase indicates phase ↑A->B or phase ↓A->B. Delay indicates delay FRR or delay FFF. Period1 refers to the period of the digital data source 1. Tablet1000 User manual Copyright © Qingdao Hantek Electronics Co., LTD...
  • Page 61: Voltage Parameters

    16. [Variance]: The average sum of the squares of the difference between the amplitude of each waveform point over the entire waveform or gating region and the mean Copyright © Qingdao Hantek Electronics Co., LTD Tablet1000 User manual...
  • Page 62: Cursor Measurement

    Drag the vertical line to the position you want to measure. ⚫ The measurement results are displayed in the upper left corner of the screen. Vertical cursor display window: Figure 10.2 Vertical cursor Tablet1000 User manual Copyright © Qingdao Hantek Electronics Co., LTD...
  • Page 63: Horizontal Cursor

    Y1 is the cursor AY, Y2 is the cursor BY. AY: Y value at cursor Y1. BY: Y value at cursor Y2. BY-AY: Vertical spacing between cursors Y1 and Y2. Copyright © Qingdao Hantek Electronics Co., LTD Tablet1000 User manual...
  • Page 64 Measure Horizontal cursor movement Use the touch screen function to adjust the cursor position by dragging gestures. For details, see Dragging section. Tablet1000 User manual Copyright © Qingdao Hantek Electronics Co., LTD...
  • Page 65: Acquisition

    This method is usually used when the sampling rate of the digital converter is higher than the storage rate of the acquisition memory. 11.2 Average times Average times: Copyright © Qingdao Hantek Electronics Co., LTD Tablet1000 User manual...
  • Page 66: Display Mode

    If the principal axis of the ellipse is in quadrants I and III, the obtained phase difference angle should be in quadrants I and IV, that is, in (0 to π/2) or (3π/2 to 2π). If the Tablet1000 User manual Copyright © Qingdao Hantek Electronics Co., LTD...
  • Page 67: Storage Depth

    11.4 Storage depth Memory depth refers to the number of waveform points an oscilloscope can store in a trigger acquisition. It reflects the storage capacity of the acquisition memory. Copyright © Qingdao Hantek Electronics Co., LTD Tablet1000 User manual...
  • Page 68: Interpolation Methods

    Interpolation methods include sine wave interpolation, step interpolation and linear interpolation. Tablet1000 User manual Copyright © Qingdao Hantek Electronics Co., LTD...
  • Page 69 Connect a straight line directly between adjacent sample points. This method is limited to reconstructing signals with straight edges, such as square waves. Sinusoidal interpolation uses curves to connect sampling points and has more universality. Copyright © Qingdao Hantek Electronics Co., LTD Tablet1000 User manual...
  • Page 70: Display

    The acquired waveforms are displayed in low brightness, and the newly acquired waveforms are displayed in normal brightness and color. Use infinite afterglow to measure noise and jitter, and capture accidental events. Tablet1000 User manual Copyright © Qingdao Hantek Electronics Co., LTD...
  • Page 71: Grid

    Grid brightness Adjust grid brightness by sliding the slider left and right with one finger. 12.5 Waveform brightness Adjust waveform brightness by sliding the slider left and right with one finger. Copyright © Qingdao Hantek Electronics Co., LTD Tablet1000 User manual...
  • Page 72: Save

    The pictures and other contents saved in the machine can be found and opened in the file browse under the main interface. Images can also be found and opened in the gallery under the main interface.See the File browse Gallery sections for details. Tablet1000 User manual Copyright © Qingdao Hantek Electronics Co., LTD...
  • Page 73: Calibration

    Exit: If oscilloscope is in the calibration process, click Exit to exit the calibration. : Close the calibration interface. If the machine completes the calibration normally and successfully, it will automatically exit the calibration interface. Copyright © Qingdao Hantek Electronics Co., LTD Tablet1000 User manual...
  • Page 74: Shortcuts

    Has been adjusted n trigger Video trigger Has been adjusted standard Vertical bandwidth DC (if GND was previously selected). For video signals, set it to Vertical coupling AC. Otherwise, no change. Tablet1000 User manual Copyright © Qingdao Hantek Electronics Co., LTD...
  • Page 75: Default Setting

    Menu or system Options, labels or knobs Default settings Acquisition Mode Normal Running state Run/stop Cursor State Type Vector Display Afterglow Mode Horizontal Window mode Single window Copyright © Qingdao Hantek Electronics Co., LTD Tablet1000 User manual...
  • Page 76: Table 15.3 Default Setting

    Table 15.3 Default setting15 The following settings do not change when the default setting is selected. ⚫ Language ⚫ Save settings ⚫ Save reference waveform ⚫ Display contrast ⚫ Calibration data Tablet1000 User manual Copyright © Qingdao Hantek Electronics Co., LTD...
  • Page 77: Multimeter

    Record the measured values in the table. The slider bar on the right side of the table allows you to view the previously saved data. Save: Click Save to save the measured value to a USB disk. Connection chart: Copyright © Qingdao Hantek Electronics Co., LTD Tablet1000 User manual...
  • Page 78: Measurement

    Connect the red and black probes to the measured point. The voltage value of the measured point will be displayed. Diode measurement ⚫ Click [DMM] on the main interface to enter the multimeter interface. ⚫ Selective diode ( Tablet1000 User manual Copyright © Qingdao Hantek Electronics Co., LTD...
  • Page 79 Connect the red and black probes to the measured point. The the voltage value of the measured point will be displayed. Note: Please correctly select the measurement gear before measuring. Copyright © Qingdao Hantek Electronics Co., LTD Tablet1000 User manual...
  • Page 80: Signal Source

    Click Symmetry to set the symmetry of the triangle wave. Click High-Z repeatedly to set the output load impedance of the waveform generator. The default value is 50Ω. Tablet1000 User manual Copyright © Qingdao Hantek Electronics Co., LTD...
  • Page 81: Set Waveform Modulation

    When the burst menu is opened, the burst is on. When the burst menu is closed, thee burst is off. Click Burst to open the burst menu. Set count and click Trigger to get the corresponding waveform state. Copyright © Qingdao Hantek Electronics Co., LTD Tablet1000 User manual...
  • Page 82: Edit Arbitrary Waveform

    Smooth drawing mode. Use the left mouse button to draw waveforms of any shape. Linear drawing mode. Click on the waveform to draw a straight line from the previous point. Tablet1000 User manual Copyright © Qingdao Hantek Electronics Co., LTD...
  • Page 83: Output Arbitrary Waveform

    Select a waveform file or draw an arbitrary waveform and click on the icon the toolbar. Then select a download location of the waveform data and download the waveform to Arb1~Arb4 in the signal source menu. Copyright © Qingdao Hantek Electronics Co., LTD Tablet1000 User manual...
  • Page 84 Plug the storage device with ARB files into the machine. Enter the signal source interface and select one among Arb1 ~ Arb4. Hold down until the message “Call out successfully” is displayed. The AFG port will output the waveform. Tablet1000 User manual Copyright © Qingdao Hantek Electronics Co., LTD...
  • Page 85: File Browse

    There are delete and rename options on the right side of the File browse interface. Delete ⚫ Select the file to delete. ⚫ Click Delete. ⚫ Pop-up: Are you sure you want to delete the file? ⚫ Select Confirm or Cancel. Copyright © Qingdao Hantek Electronics Co., LTD Tablet1000 User manual...
  • Page 86 Click Rename. ⚫ The keyboard will pop up. ⚫ Set the name. ⚫ Click Enter to complete the renaming. For specific operations, please refer to the section of Setting parameters. Tablet1000 User manual Copyright © Qingdao Hantek Electronics Co., LTD...
  • Page 87: Settings

    Two languages: English or Chinese. System information Display the machine model, serial number, software version, hardware version, and firmware version. Voice Use the slider to adjust the volume. Change the boot music as prompted. Copyright © Qingdao Hantek Electronics Co., LTD Tablet1000 User manual...
  • Page 88 Follow these steps to perform USB sharing: Connect the machine to the computer by USB. Click USB sharing and click the USB sharing tab in the middle of the screen. Tablet1000 User manual Copyright © Qingdao Hantek Electronics Co., LTD...
  • Page 89 Settings The computer pops up a removable disk and opens it to read files stored in the machine (mini_pic、image、ref、csv、setup、wave、mask). Click ON/OFF to exit USB sharing. Copyright © Qingdao Hantek Electronics Co., LTD Tablet1000 User manual...
  • Page 90: Gallery

    If you delete an image in the middle, the next saved image will be saved in the same place as the previous deleted image, and the images saved later will continue in the previous order. Tablet1000 User manual Copyright © Qingdao Hantek Electronics Co., LTD...
  • Page 91: Remote Control

    Double-click the IO software, and you can view the connected device information displayed under My Instrument-USB. Click "Interactive IO" and send an instruction, then the computer is in communication with the oscilloscope. Copyright © Qingdao Hantek Electronics Co., LTD Tablet1000 User manual...
  • Page 92 Remote control Double-click to open the upper computer software. If you see the connection information in the lower left corner of the interface, it indicates the upper computer has been connected. Tablet1000 User manual Copyright © Qingdao Hantek Electronics Co., LTD...
  • Page 93: Fault Handling

    Check whether the power button is pressed firmly. ⚫ After the above examination, restart the instrument. ⚫ If you still cannot use this product properly, please contact Hantek. After the oscilloscope is turned on, no waveform is displayed on the input signal screen. ⚫...
  • Page 94: Appendix

    2-channel oscilloscope TO1152C multimeter Sa/S, 200MHz, 2-channel oscilloscope TO1202C multimeter Sa/S, 250MHz, 2-channel oscilloscope TO1252C multimeter Sa/S, 150MHz, 2-channel oscilloscope TO1152D multimeter + signal source Sa/S, 200MHz, 2-channel oscilloscope TO1202D Tablet1000 User manual Copyright © Qingdao Hantek Electronics Co., LTD...
  • Page 95 + signal source Sa/S, 200MHz, 4-channel oscilloscope TO1204D multimeter + signal source Sa/S, 250MHz, 4-channel oscilloscope TO1254D multimeter + signal source Sa/S, 150MHz, 4-channel oscilloscope TO1154AUTO multimeter + signal source Copyright © Qingdao Hantek Electronics Co., LTD Tablet1000 User manual...
  • Page 96: Appendix B: Warranty Summary

    During the warranty period, if the product is proved to be defective, Hantek will repair or replace the product free of charge. Please refer to the description on Hantek official website for detailed warranty regulations.
  • Page 97 Appendix liable for indirect, special or consequential damages. Copyright © Qingdao Hantek Electronics Co., LTD Tablet1000 User manual...
  • Page 98 Addr: #35 Building, No. 780 Baoyuan Road, High-tech Zone, Qingdao, Shandong, China 266114 Switchboard: 400-036-7077 Email: service@hantek.com Tel: (0086)532-55678770 & 55678772 & 55678773 Zip code: 266114 Website: www.hantek.com Qingdao Hantek Electronics Co., LTD...