Troubleshooting
Message
Caused by
Problem with the test
strip
The blood sample was
applied before the (blood
insertion picture) symbol
appeared on the display
(refer p21).
The ambient temperature
is too low
38
What to do
Insert a new test strip. If Er4
message persists, please
contact
your
authorized
representative or customer
support.
Repeat the test with new test
strip. Apply blood only after the
(blood insertion picture) symbol
appears on the display.
Place the meter within the
operating temperature range for
more than 10 minutes and
retest.
Message
Caused by
The ambient temperature
is too High
The test result is lower
than 10 mg/dl
(0.6 mmol/L)
The test result is higher
than 600mg/dl
(33.3 mmol/L)
Troubleshooting
What to do
Place the meter within the
operating temperature range for
more than 10 minutes and
retest.
Check the accuracy of your test
strip by performing a control
solution test. After, re-test with
blood sample two or three
times. When "Lo" message
persists, please consult a doctor
for assistance.
Check the accuracy of your test
strip by performing a control
solution test. After, re-test with
blood sample two or three
times. When "HI" message
persists, please consult a doctor
for assistance.