Checking the system with
the
Testing procedure
1. Prior to use, shake the bottle gently several times.
2. Invert bottle and discard the first drop of control solution.
3. Wipe the dispenser tip clean.
4. Test strip is inserted fully into the test slot of the EASYGLUCO Auto-
meter.
coding
TM
5. Be sure the code number appearing on the LCD window matches the test
strip code on the strip vial.
6. When the strip sign blinks, press
control solution measuring process appears. Effect: The result is marked.
7. Hold the control vial at an angle with the tip near the end of the test strip.
Gently squeeze the vial and form a small drop. Pick up the meter and touch
the end of the test strip to the hanging drop. Be sure to hold the test strip to
the drop until the "beep" sounds.
8. After the "beep" sound, wipe the rest of the drop with tissue or cotton gauze.
If the drop is taken back into the control vial, it is possible to contaminate the
whole vial of control solution!
9. The displayed test result should be within the range printed on the test strip
vial.
Second application tip : Place a drop on clean aluminum foil, glass or plastic plate and
touch the end of the test strip to that drop. This method is more favorable, in preventing
contamination to the whole vial of control solution
IMPORTANT NOTE : If the
EASYGLUCO Auto-coding
continuously, falls outside the range printed on the vial, the
system may not be functioning properly. DO NOT use the system to test your
coding
TM
blood until you get a control solution test result that falls within the range. If you continue
to have problems, call your Customer Support and Service.
In the User ID option is activated, the control solution test result is stored, but can only be
recalled after deactivating the User ID option.
18
control solution
button once. The symbol
<s
or
Control Solution test results,
TM
EASYGLUCO Auto-
Before testing, be sure to read this section and test strip package insert found
in the test strip box carefully. Make sure you have all the necessary items to
begin the test:
for
Caution
To reduce risk of infection :
1. The lancing device and sterile lancets should NOT be shared with others.
2. Always use a new, sterile lancet and a new strip
3. Practice using the Lancing Device and become accustomed with its use.
4. Lancets and strips are for single use only.
5. Wash your hands with warm, clean water and soap before testing.
6. Avoid getting hand lotion, oils, dirt or debris on the lancets or on the lancing
device.
7. Dry your hands completely before testing.
8. Sterile lancets are for single use only. Do not reuse.
9. Sterile lancets is used with a spring-operated lancing device
10. Sterile lancets sterilized using gamma irradiation. After opening, exposed parts
should not be touched or put on the floor to prevent contamination.
Testing Your Blood - Preparation
1. EASYGLUCO Auto-coding
2. EASYGLUCO Auto-coding
Strip
3. Lancing Device
4. Sterile lancets
Meter
TM
Test
TM
19