WATER-COOLED CONNECTIONS
Increase coolant flow by an additional 1.0 GPM (3.8 lpm). Do not exceed a coolant pressure of 60
PSI (4.1 bar).
Air-condition the room or the enclosure containing the laser.
Install a dehumidifier to reduce the humidity of the enclosure containing the laser.
Table 1-2 provides dew point temperatures for a range of air temperature and relative humidity values.
Remember that the laser's coolant temperature must be set above the dew point temperatures shown in
the chart but should not exceed 22 °C (72 °F).
Table 1-2 Dew point temperatures