Table 4 Recommended separation distances between portable and mobile RF
communications equipment and the S/5™ CM (1
Recommended separation distances between portable and mobile RF communications
equipment and the S/5™ CM (1 .
The S/5™ CM (1 is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the S/5™ CM (1 can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equipment (transmitters)
and the S/5™ CM (1 as recommended below, according to the maximum output power of the communications
equipment.
Rated maximum
output power of
transmitter
W
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in
meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2
These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
(1
S/5™ CM equipped with M-NESTPR or M-NE12STPR, M-ENTROPY, M-BIS, M-MINIC and N-DIS.
(2
For M-BIS.
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
d = 1.2 P
0.12
0.38
1.2
3.8
12
APPENDIX B, EMC tables, S/5 Compact Monitors
80 MHz to 800 MHz
d = 1.2 P
(2
d = 3.5 P
0.12
(2
0.35
0.38
(2
1.1
1.2
(2
3.5
3.8
(2
11
12
(2
35
800 MHz to 2.5 GHz
d = 2.3 P
(2
d = 7.0 P
0.23
(2
0.70
0.73
(2
2.2
2.3
(2
7.0
7.3
(2
22
23
(2
70
Document No. M1031508
B-5(6)