Table 3 Guidance and manufacturer's declaration – electromagnetic immunity
Guidance and manufacturer's declaration – electromagnetic immunity
(1
The S/5™ CM
is intended for use in the electromagnetic environment specified below. The customer or the user of the
(1
S/5™ CM
should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to 80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2.5 GHz
3 V/m
80 MHz to 2.5 GHz
NOTE 1
At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
APPENDIX B, EMC tables, S/5 Compact Monitors
Compliance
Electromagnetic environment - guidance
level
Portable and mobile RF communications equipment should
be used no closer to any part of the S/5™ CM
cables, than the recommended separation distance
calculated from the equation applicable to the frequency of
the transmitter.
Recommended separation distance
d = 1.2 P
3 Vrms
d = 1.2 P 80 MHz to 800 MHz
3 V/m
d = 2.3 P 800 MHz to 2.5 GHz
(2
1 V/m
d = 3.5 P 80 MHz to 800 MHz
d = 7.0 P 800 MHz to 2.5 GHz
where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in metres (m).
Field strengths from fixed RF transmitters, as determined by
an electromagnetic site survey,
compliance level in each frequency range.
Interference may occur in the vicinity of equipment marked
with the following symbol:
(1
, including
a
should be less than the
b
B-3(6)
Document No. M1031508