4-38
Maintenance: Alarm Test
10. Access Install Options Screen 2.
11. Select Store Current to Hospital. Exit the service mode by selecting Restart at the
bottom of the screen.
12. Set the switches on the Model 325 Input Simulator according to Table 8.
13. Using the simulator's Manual Adjustment knob, input an MECG signal of 119
bpm as indicated on the monitor. Verify that there is no alarm tone sounding
from the monitor's rear panel speaker.
14. Using the simulator's Manual Adjustment knob, increase the MECG rate to 120
bpm. Again, verify that there is no alarm tone sounding from the rear panel
speaker.
15. Using the simulator's Manual Adjustment knob, increase the MECG rate to 121
bpm. Verify the following on the monitor:
The following alarm tone is emitted from the rear panel speaker:
alternating high/low tones until the alarm condition is removed (following
steps.)
The MECG value flashes.
16. Depress the monitor's front panel Alarm Silence button and verify the following:
The alarm tone is silenced.
The ALARM SILENCE X:XX message box appears on the screen and a
countdown is started.
Table 8. Alarm Test Simulator Settings
Section
FECG/MECG
GENERAL
17. Wait the user-specified re-alarm time and verify the following:
The alarm tone is once again emitted from the rear panel speaker.
18. Use the simulator's Manual Adjustment knob to decrease the MECG rate to 120
bpm. Verify the following on the monitor:
The alarm tone is silenced.
The MECG value no longer flashes.
After 10 seconds, the two above conditions are still true.
19. Using the simulator's Manual Adjustment knob, input an MECG signal of 61
bpm. Verify that there is no alarm tone sounding from the rear panel speaker.
20. Using the simulator's Manual Adjustment knob, decrease the MECG rate to 60
bpm. Again, verify that there is no alarm tone sounding from the rear panel
speaker.
250cx Series Maternal/Fetal Monitor
2036947-001
Switch
Main
Rate
Mode
QRS Amplitude
QRS Polarity
Pattern Memory
Setting
Rate
Manual
MECG
500 µV
+
Off
Revision C