Type
Alarm Defaults
Audio: on
Volume: 5
Limits: High = 120 bpm, Low = 50
bpm
Re-alarm: 120 sec
MHR/P limit (high or low) actively
being violated.
MHR/P
1
The tachycardia response time is < 8
seconds.
Asystole.
MECG leads off.
1
There is an MECG re-alarm.
Alarm Defaults
Audio: on
Volume: 5
Limits: High = 100%, Low = 95%
Re-alarm: 120 sec
MSpO
being violated. Issued after about 8
seconds.
MSpO
2
Malfunction with MSpO
MSpO
disconnected from monitor; sensor
assembly disconnected from
intermediate cable; or sensor or
cable has a broken wire.
B-4
Alarms Summary:
Table B-1. Summary of 250cx Series Alarms
Condition
limit (high or low) actively
2
circuitry.
2
intermediate cable
2
NOTE: For further information on Alarms, refer to Chapter 11 in the
Corometrics 250cx Series Operator's Manual.
250cx Series Maternal/Fetal Monitor
2036947-001
Visible Advisory
MHR/P numeric flashes.
Flashing dashes "– – –" in place
of MHR/P numeric.
Flashing dashes "– – –" in place
of MHR/P numeric and MECG
LEADS OFF message displays
underneath.
MSpO
numeric flashes. MSpO
2
value and pulse rate print on the
strip chart.
COMM or REPAIR message
displays in MSpO
area.
2
Dashes "– – –" in place of MSpO
numeric.
Audible Advisory
Alternating high/low tones (if
audio enabled).
Alternating high/low tones (if
audio enabled).
Alternating high/low tones (if
audio enabled).
Alternating high/low tones (if
2
audio enabled).
Alternating high/low tones (if
audio enabled).
Alternating high/low tones (if
2
audio enabled).
Revision C