4-18
Maintenance: Self-Test Routine
in the FHR2 area, with both mode titles displaying Test.
The monitor adds 50 mmHg to the present pressure input level and displays
this value in the UA display area; the mode title displays Test.
If MECG is the selected waveform, a simulated waveform appears on the
display.
NOTE: The monitor will add 50 mmHg to raw pressure data. In other words, the
value is always referenced to 0 mmHg regardless of any UA referencing
attempt.
NOTE: The recorder returns to its original on, off, or maternal-only mode state
from when the Test button was depressed.
NOTE: To disable the test, depress the Test button or open the recorder door.
NOTE: If the simulated fetal heart rate trends do not appear in the correct positions
on the strip chart recorder paper, ensure the monitor's paper scale (30-240
bpm or 50-210 bpm) setting matches the type of paper being used, i.e., 30
bpm/cm or 20 bpm/cm. (Refer to"Install Options Screens" on page 3-14.)
250cx Series Maternal/Fetal Monitor
2036947-001
250cx Series Self-Test
Revision C